Features and Functions 3<br />

ABORt<br />

DISPlay:TEXT ''<br />

OUTPut[:STATe] {OFF|0|ON|1}, (@)<br />

ROUTe:CLOSe (@)<br />

ROUTe:CLOSe:EXCLusive (@)<br />

ROUTe:MODule:WAIT {1-8|SLOT1-SLOT8|ALL}<br />

ROUTe:OPEN (@)<br />

ROUTe:OPEN:ABUS [{1-4|ABUS1-ABUS4|ALL}]<br />

ROUTe:OPEN:ALL [{1-8|SLOT1-SLOT8|ALL}]<br />

ROUTe:SEQuence:TRIGger[:IMMediate] <br />

[SENSe:]TOTalize:CLEar:IMMediate (@)<br />

SOURce:CURRent[:LEVel] {|MIN|MAX|DEF}, (@)<br />

SOURce:DIGital:DATA[:{BYTE|1|WORD|2|LWORd|4}] ,(@)<br />

SOURce:DIGital:DATA:BIT {0|1}, , (@)<br />

SOURce:FUNCtion:TRIGger:IMMediate (@)<br />

SOURce:VOLTage[:LEVel] {|MIN|MAX|DEF} , (@)<br />

SYSTem:BEEPer<br />

SYSTem:DELay[:IMMediate] <br />

Sequences can be defined from the remote interface only. You can,<br />

however, review, execute, and delete sequences from the front panel.<br />

When a sequence is defined, the specified commands are checked for<br />

proper syntax and absolute parameter range limits. If an error is<br />

detected during compilation, the entire sequence will be discarded.<br />

During compilation, the sequence commands do not have to be valid for<br />

the current instrument configuration; this allows you to define<br />

sequences without regard to compatibility with the current set of<br />

installed modules. More extensive error checking, such as channel range<br />

expansion and validation, is performed when the sequence is executed.<br />

If you define a sequence with a name already in use by another<br />

sequence, the new definition will overwrite the previous definition <br />

(no error is generated).<br />

A sequence name can contain up to 30 characters. The first character<br />

must be a letter (A- Z), but the remaining 29 characters can be letters,<br />

numbers (0- 9), or an underscore ( _ ). Blank spaces are not allowed.<br />

When stored in memory, the user- defined sequence names are<br />

converted to all uppercase letters. For example, when stored “MySeq_1”<br />

is converted to “MYSEQ_1”.<br />