Features and Functions 3<br />
ABORt<br />
DISPlay:TEXT ''<br />
OUTPut[:STATe] {OFF|0|ON|1}, (@)<br />
ROUTe:CLOSe (@)<br />
ROUTe:CLOSe:EXCLusive (@)<br />
ROUTe:MODule:WAIT {1-8|SLOT1-SLOT8|ALL}<br />
ROUTe:OPEN (@)<br />
ROUTe:OPEN:ABUS [{1-4|ABUS1-ABUS4|ALL}]<br />
ROUTe:OPEN:ALL [{1-8|SLOT1-SLOT8|ALL}]<br />
ROUTe:SEQuence:TRIGger[:IMMediate] <br />
[SENSe:]TOTalize:CLEar:IMMediate (@)<br />
SOURce:CURRent[:LEVel] {|MIN|MAX|DEF}, (@)<br />
SOURce:DIGital:DATA[:{BYTE|1|WORD|2|LWORd|4}] ,(@)<br />
SOURce:DIGital:DATA:BIT {0|1}, , (@)<br />
SOURce:FUNCtion:TRIGger:IMMediate (@)<br />
SOURce:VOLTage[:LEVel] {|MIN|MAX|DEF} , (@)<br />
SYSTem:BEEPer<br />
SYSTem:DELay[:IMMediate] <br />
Sequences can be defined from the remote interface only. You can,<br />
however, review, execute, and delete sequences from the front panel.<br />
When a sequence is defined, the specified commands are checked for<br />
proper syntax and absolute parameter range limits. If an error is<br />
detected during compilation, the entire sequence will be discarded.<br />
During compilation, the sequence commands do not have to be valid for<br />
the current instrument configuration; this allows you to define<br />
sequences without regard to compatibility with the current set of<br />
installed modules. More extensive error checking, such as channel range<br />
expansion and validation, is performed when the sequence is executed.<br />
If you define a sequence with a name already in use by another<br />
sequence, the new definition will overwrite the previous definition <br />
(no error is generated).<br />
A sequence name can contain up to 30 characters. The first character<br />
must be a letter (A- Z), but the remaining 29 characters can be letters,<br />
numbers (0- 9), or an underscore ( _ ). Blank spaces are not allowed.<br />
When stored in memory, the user- defined sequence names are<br />
converted to all uppercase letters. For example, when stored “MySeq_1”<br />
is converted to “MYSEQ_1”.<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 141