Features and Functions 3<br />

You can specify the number of times the instrument will sweep through<br />

the scan list. When the specified number of sweeps have occurred, the<br />

scan stops. For more information, refer to “Sweep Count” on page 117.<br />

An externally- controlled scan can also include a read of a digital port<br />

or a read of the totalizer count on the digital modules. When the<br />

channel advance reaches the first digital channel, the instrument scans<br />

through all of the digital channels in that slot (only one channel<br />

advance signal is required).<br />

You can configure the list of channels for 4- wire external scanning<br />

without the internal DMM. When enabled, the instrument automatically<br />

pairs channel n in Bank 1 with channel n+20 in Bank 2 (34921A,<br />

34923A, 34925A) or n+35 (34922A, 34924A) to provide the source and<br />

sense connections. For example, make the source connections to the HI<br />

and LO terminals on channel 2 in Bank 1 and the sense connections to<br />

the HI and LO terminals on channel 22 (or 37) in Bank 2.<br />

Front Panel Operation: To select the channel advance source, choose the<br />

following items.<br />

Scan (Configure) > ADVANCE CHANNEL > AUTO|EXT|MANUAL<br />

To initiate the scan and store all readings in memory, press the <br />

Scan (<strong>Measure</strong>) key.<br />

To configure the instrument for 4- wire external scanning, choose the<br />

following menu item.<br />

Channel (Configure) > FOUR WIRE > OFF|ON<br />

Remote Interface Operation: The following program segment configures the<br />

instrument for an externally- controlled scan. <br />

INST:DMM OFF<br />

ROUT:SCAN (@1001:1020)<br />

TRIG:SOUR IMM<br />

TRIG:COUN 5<br />

ROUT:CHAN:ADV:SOUR EXT<br />

INIT<br />

Disable internal DMM<br />

Configure scan list<br />

Set trigger source<br />

Set trigger count<br />

Set channel advance source<br />

Initiate the scan<br />

To configure the instrument for 4- wire external scanning, send the<br />

following command.<br />

ROUTe:CHANnel:FWIRe {OFF|ON}, (@)<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 131