Totalizer<br />

Introduction to the <strong>34980A</strong> 1<br />

The digital modules have 32- bit totalizer input channels for counting<br />

pulses and measuring frequency. You can manually read the totalizer<br />

count or you can configure a scan to read the count.<br />

32 Bits<br />

Totalizer<br />

Count +<br />

Count -<br />

You can configure the totalizer to count on the rising edge or falling edge<br />

of the input signal. The minimum rise/fall time for input signals to the<br />

totalizer is 5 S.<br />

The maximum count is 4,294,967,295 (2 32 - 1). The count rolls over to "0"<br />

after reaching the maximum allowed value.<br />

You can configure the totalizer to read without affecting the count or<br />

reset the count to zero without losing any counts.<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 17<br />

Gate<br />

Gate