3 Features and Functions<br />

Front Panel Operation: To select the thermocouple function on the active<br />

channel, choose the following items.<br />

DMM or Channel (Configure) > TEMPERATURE > PROBE TYPE > THERMOCOUPLE<br />

Then, use the knob to select the thermocouple type from the list.<br />

THERMOCOUPLE TYPE > B|E|J|K|N|R|S|T<br />

If desired, you can enable the thermocouple check feature on the active<br />

channel (opens are reported as “OPEN T/C”).<br />

T/C CHECK > OFF|ON<br />

To select the reference junction source for the active channel, choose one<br />

of the following items.<br />

REFERENCE > FIXED|EXT|INT<br />

For an external reference, configure an RTD or thermistor as the external<br />

reference channel.<br />

Channel (Configure) > TEMPERATURE > PROBE TYPE > RTD > . . . USE AS EXT REF?<br />

Remote Interface Operation: You can use the CONFigure or MEASure?<br />

command to select the probe type and thermocouple type. For example,<br />

the following command configures channel 3001 for a J- type thermocouple<br />

measurement.<br />

CONF:TEMP TC,J,(@3001)<br />

You can also use the SENSe command to select the probe type and<br />

thermocouple type. For example, the following command configures<br />

channel 2003 for a J- type thermocouple measurement.<br />

SENS:TEMP:TRAN:TC:TYPE J,(@2003)<br />

The following commands use the SENSe command to set a fixed reference<br />

junction temperature of 40 degrees (always in °C) on channel 2003.<br />

SENS:TEMP:TRAN:TC:RJUN:TYPE,(@2003)<br />

SENS:TEMP:TRAN:TC:RJUN 40,(@2003)<br />

The following command enables the thermocouple check feature on the<br />

specified channel (opens are reported as “+9.90000000E+37”).<br />

RTD <strong>Measure</strong>ments<br />

SENS:TEMP:TRAN:TC:CHECK ON,(@2003)<br />

The instrument supports RTDs with = 0.00385 (DIN/IEC 751) using<br />

ITS- 90 software conversions or = 0.00391 using IPTS- 68 software<br />

conversions. The default is = 0.00385.<br />