To Build a Scan List From the Remote Interface<br />
Features and Functions 3<br />
Use the ROUTe:SCAN command to define the list of channels in the scan<br />
list. To determine what channels are currently in the scan list, use the<br />
ROUTe:SCAN? query command.<br />
To add channels to the present scan list, use the ROUTe:SCAN:ADD<br />
command. To remove channels from the present scan list, use the<br />
ROUTe:SCAN:REMove command.<br />
To remove all channels from the scan list, send “ROUT:SCAN (@)”.<br />
To initiate a scan, use the INITiate or READ? command. <br />
<strong>Measure</strong>ments are stored in memory. Each time you initiate a new scan,<br />
the instrument will clear the previous set of readings from memory.<br />
To stop a scan in progress, use the ABORt command.<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 111