To Build a Scan List From the Remote Interface<br />

Features and Functions 3<br />

Use the ROUTe:SCAN command to define the list of channels in the scan<br />

list. To determine what channels are currently in the scan list, use the<br />

ROUTe:SCAN? query command.<br />

To add channels to the present scan list, use the ROUTe:SCAN:ADD<br />

command. To remove channels from the present scan list, use the<br />

ROUTe:SCAN:REMove command.<br />

To remove all channels from the scan list, send “ROUT:SCAN (@)”.<br />

To initiate a scan, use the INITiate or READ? command. <br />

<strong>Measure</strong>ments are stored in memory. Each time you initiate a new scan,<br />

the instrument will clear the previous set of readings from memory.<br />

To stop a scan in progress, use the ABORt command.<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 111