3 Features and Functions<br />

Trigger Delay<br />

In some applications, you want to allow the input to settle before taking a<br />

reading or for pacing a burst of readings. You can add a trigger delay,<br />

which adds a delay between the trigger signal and the first sample taken<br />

by the internal DMM (not used in Scanning Mode). The programmed<br />

trigger delay overrides the default trigger delay that the instrument<br />

automatically adds to the measurement.<br />

Trigger 1 Sample Count Trigger 2 Sample Count<br />

Trigger delay<br />

Trigger Delay<br />

(0 to 3600 seconds)<br />

The default trigger delay is Automatic (see “Automatic Trigger<br />

Delays” on page 91); the instrument determines the delay based on<br />

function, range, and integration time.<br />

If you specify a trigger delay other than Automatic, that same delay is<br />

used for all functions and ranges.<br />

If you have configured the instrument to take more than one reading<br />

per trigger (sample count > 1), the specified trigger delay is inserted<br />

between the trigger and the first reading in the sample burst.<br />

The CONFigure and MEASure? commands set the trigger delay to<br />

Automatic.<br />

The instrument selects an automatic trigger delay after a Factory Reset<br />

(*RST command). An Instrument Preset (SYSTem:PRESet command) or<br />

Card Reset (SYSTem:CPON command) does not change the setting.<br />

90 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide<br />