Resistance <strong>Measure</strong>ment Configuration<br />

Features and Functions 3<br />

This section contains information to help you configure the instrument <br />

for making resistance measurements. Use the 2- wire method for ease of<br />

wiring and higher density or use the 4- wire method for improved<br />

measurement accuracy. The measurement ranges shown below.<br />

Offset Compensation<br />

100 1 k 10 k 100 k 1 M 10 M 100 M Autorange<br />

Offset compensation removes the effects of any dc voltages in the circuit<br />

being measured. The technique involves taking the difference between two<br />

resistance measurements on the specified channels, one with the current<br />

source turned on and one with the current source turned off.<br />

Applies only to 2- wire and 4- wire resistance measurements on the <br />

100, 1 k, and 10 k ranges.<br />

Four- wire measurements are not allowed on the multiplexer modules<br />

configured for the 1- wire (single ended) mode (see page 95).<br />

For 4- wire resistance measurements, the instrument automatically pairs<br />

channel n in Bank 1 with channel n+20 in Bank 2 (34921A, 34923A,<br />

34925A) or n+35 (34922A, 34924A) to provide the source and sense<br />

connections. For example, make the source connections to the HI and<br />

LO terminals on channel 2 in Bank 1 and the sense connections to the<br />

HI and LO terminals on channel 22 (or 37) in Bank 2.<br />

The CONFigure and MEASure? commands automatically disable offset<br />

compensation.<br />

The instrument disables offset compensation after a Factory Reset<br />

(*RST command). An Instrument Preset (SYSTem:PRESet command) or<br />

Card Reset (SYSTem:CPON command) does not change the setting.<br />

Front Panel Operation: DMM or Channel (Configure) > OFFSET COMP<br />

Remote Interface Operation:<br />

[SENSe:]FRESistance:OCOMpensated {OFF|ON} [,(@)]<br />

[SENSe:]RESistance:OCOMpensated {OFF|ON} [,(@)]<br />

If you omit the optional parameter, the command applies to the<br />

internal DMM. For 4- wire measurements, specify the paired channel in<br />

Bank 1 (source) as the channel (channels in Bank 2 are not<br />

allowed in the ).<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 103