3 Features and Functions<br />

ABus1 HI<br />

ABus2 HI<br />

ABus3 HI<br />

ABus4 HI<br />

Analog Bus Connector<br />

9<br />

6<br />

5<br />

1<br />

ABus1 LO<br />

ABus2 LO<br />

ABus3 LO<br />

ABus4 LO<br />

<strong>34980A</strong><br />

Chan Adv In<br />

GND<br />

Chan Closed Out<br />

Ext Trig Connector<br />

For an externally- controlled scan, you must either remove the internal<br />

DMM from the <strong>34980A</strong> or disable it (see “Internal DMM Disable” on<br />

page 153). Since the <strong>34980A</strong>’s internal DMM is not used, readings from<br />

multiplexer channels are stored in the external DMM’s memory.<br />

In this configuration, you must set up a scan list to include all desired<br />

multiplexer or digital channels. Channels which are not in the list are<br />

skipped during the scan. By default, the instrument scans the list of<br />

channels in ascending order from slot 1 through slot 8 (channels are<br />

reordered as needed).<br />

You can configure the event or action that controls the onset of each<br />

sweep through the scan list (a sweep is one pass through the scan list).<br />

The selected source is used for all channels in the scan list. For more<br />

information, refer to “Scan Trigger Source” on page 112.<br />

You can configure the event or action that notifies the <strong>34980A</strong> to<br />

advance to the next channel in the scan list. Note that the Channel<br />

Advance source shares the same sources as the scan trigger. However,<br />

an error is generated if you attempt to set the channel advance source<br />

to the same source (other than IMMediate) used for the scan trigger.<br />

130 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide<br />

6<br />

9<br />

1<br />

5<br />

External DMM<br />

VM Complete Out Ext Trig In