3 Features and Functions<br />
Sequences<br />
This section gives information on defining and executing a sequence, <br />
which is a compiled series of SCPI commands stored in non- volatile<br />
memory and identified by a user- defined name. Sequences can be used in<br />
a variety of applications, such as creating a signal path from a<br />
device- under- test to a measurement device or sequencing relays in a<br />
specified order. You can also uses sequences in conjunction with other<br />
operations to configure and synchronize complex measurements without<br />
having to send the routing commands each time.<br />
The following tables summarizes the commands used to define, execute,<br />
and manage sequences. For more information, see the <strong>Agilent</strong> <strong>34980A</strong><br />
Programmer’s Reference Help file.<br />
Defining a Sequence<br />
Sequence Definition<br />
ROUTe:SEQuence:DEFine , ""<br />
ROUTe:SEQuence:DEFine? <br />
Sequence Execution<br />
ROUTe:SEQuence:ABORT<br />
ROUTe:SEQuence:BUSY?<br />
ROUTe:SEQuence:RUNNing:NAME?<br />
ROUTe:SEQuence:TRIGger[:IMMediate]<br />
ROUTe:SEQuence:WAIT<br />
Sequence Management<br />
ROUTe:SEQuence:CATalog?<br />
ROUTe:SEQuence:DELete:ALL<br />
ROUTe:SEQuence:DELete[:NAME] <br />
Alarm Limits<br />
OUTPut:ALARm{1-4}:SEQuence?<br />
ROUTe:SEQuence:TRIGger:SOURce , <br />
ROUTe:SEQuence:TRIGger:SOURce? <br />
Defines a sequence.<br />
Returns sequence definition.<br />
Terminates currently-running sequence.<br />
Returns “1” if sequence is executing (busy).<br />
Returns name of currently-running sequence.<br />
Executes specified sequence.<br />
Blocks until sequence has completed.<br />
Returns list of defined sequence names.<br />
Deletes all sequences from memory.<br />
Deletes specified sequence from memory.<br />
Returns sequence associated with alarm.<br />
Assigns trigger source to sequence.<br />
Returns trigger source currently selected.<br />
A sequence defines a series of SCPI commands with an associated name.<br />
When the sequence is first defined, the commands are compiled and then<br />
stored in a compressed format in non- volatile memory. The following SCPI<br />
commands are allowed in a sequence definition (all other commands will<br />
generate an error).<br />