Features and Functions 3<br />
During a scan, the instrument automatically adds a time stamp to all<br />
readings and stores them in memory. Each reading is stored with<br />
measurement units, time stamp, channel number, and alarm status<br />
information. From the remote interface, you can specify which information<br />
you want returned with the readings. The examples below show a reading<br />
in relative and absolute format with all fields enabled.<br />
Relative Format (Default): <br />
1<br />
2<br />
Absolute Format:<br />
1<br />
2<br />
3<br />
2.61950000E+01 C,000000000.017,1003,2<br />
1 2 3 4<br />
Reading with units (26.195 °C)<br />
Time since start of scan (17 ms)<br />
Reading with units (26.195 °C)<br />
Date (November 21, 2004)<br />
Time of day (3:30:23.000 PM)<br />
The reading format applies to all readings being removed from the<br />
instrument from a scan; you cannot set the format on a per- channel<br />
basis.<br />
The CONFigure and MEASure? commands automatically turn off the<br />
units, time, channel, and alarm information.<br />
The format settings are stored in volatile memory and will be lost when<br />
power is turned off or after a Factory Reset (*RST command).<br />
Remote Interface Operation: Use the following commands to select the<br />
reading format.<br />
FORMat:READing:ALARm ON<br />
FORMat:READing:CHANnel ON<br />
FORMat:READing:TIME ON<br />
FORMat:READing:TIME:TYPE {ABSolute|RELative}<br />
FORMat:READing:UNIT ON<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 123<br />
3<br />
4<br />
4<br />
5<br />
Channel number<br />
Alarm limit threshold crossed<br />
(0 = No Alarm, 1 = LO, 2 = HI)<br />
2.61950000E+01 C,2004,11,21,15,30,23.000,1003,2<br />
1 2<br />
3 4 5<br />
Channel number<br />
Alarm limit threshold crossed<br />
(0 = No Alarm, 1 = LO, 2 = HI)