3 Features and Functions<br />
AC Low Frequency Filter<br />
The instrument uses three different ac filters which enable you to either<br />
optimize low- frequency accuracy or achieve faster ac settling times.<br />
The instrument selects the slow (3 Hz), medium (20 Hz), or fast (300 Hz)<br />
filter based on the input frequency that you specify for the selected<br />
channels or the internal DMM.<br />
Applies to ac voltage and ac current measurements only.<br />
AC Filter Settings for Voltage <strong>Measure</strong>ments:<br />
Input Frequency Default Settling Delay Minimum Settling Delay<br />
3 Hz to 300 kHz (Slow)<br />
20 Hz to 300 kHz (Medium)<br />
200 Hz to 300 kHz (Fast)<br />
7 seconds / reading<br />
1 second / reading<br />
0.12 seconds / reading<br />
1.5 seconds<br />
200 ms<br />
20 ms<br />
The CONFigure and MEASure? commands automatically select the 20 Hz<br />
(medium) filter.<br />
The instrument selects the default 20 Hz (medium) filter after a Factory<br />
Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or<br />
Card Reset (SYSTem:CPON command) does not change the setting.<br />
Front Panel Operation: DMM or Channel (Configure) > AC FILTER<br />
Remote Interface Operation: Specify the lowest frequency expected in <br />
the input signal on the specified channels. The instrument selects the<br />
appropriate filter based on the frequency you specify (see table above).<br />
[SENSe:]VOLTage:AC:BANDwidth {3|20|200} [,(@)]<br />
If you omit the optional parameter, the command applies to the<br />
internal DMM.<br />
102 <strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide