Features and Functions 3<br />
The CONFigure and MEASure? commands automatically set the scan<br />
interval to immediate (0 seconds) and the scan count to 1 sweep.<br />
The instrument sets the scan interval to immediate (0 seconds) after a<br />
Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet<br />
command) or Card Reset (SYSTem:CPON command) does not change <br />
the setting.<br />
Front Panel Operation: Scan (Configure) > INTERVAL > SCAN INTERVAL<br />
To initiate the scan and store all readings in memory, press the <br />
Scan (<strong>Measure</strong>) key. Between scan sweeps, “WAITING FOR TRIG” will be<br />
displayed on the front panel.<br />
Note: To stop a scan, press and hold the Scan (<strong>Measure</strong>) key.<br />
Remote Interface Operation: The following program segment configures the<br />
instrument for an interval scan.<br />
TRIG:SOURCE TIMER<br />
TRIG:TIMER 5<br />
TRIG:COUNT 2<br />
INIT<br />
Note: To stop a scan, send the ABORt command.<br />
Manual Scanning<br />
Select interval time mode<br />
Set the scan interval to 5 seconds<br />
Sweep the scan list 2 times<br />
Initiate the scan<br />
In this configuration, the instrument waits for either a front- panel key<br />
press or a remote interface command before sweeping through the<br />
scan list.<br />
All readings from the scan are stored in volatile memory. <br />
Readings accumulate in memory until the scan is terminated <br />
(until the trigger count is reached or until you abort the scan).<br />
You can specify a trigger count which sets the number of front- panel<br />
key presses or scan trigger commands that will be accepted before<br />
terminating the scan. See “Trigger Count” on page 117 for more<br />
information.<br />
Mx+B scaling and alarm limits are applied to measurements during a<br />
manual scanning operation and all data is stored in volatile memory.<br />
Front Panel Operation: Scan (Configure) > INTERVAL > MANUAL<br />
To initiate the scan and store all readings in memory, press the <br />
Scan (<strong>Measure</strong>) key.<br />
Note: To stop a scan, press and hold the Scan (<strong>Measure</strong>) key.<br />
<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 113