Trigger Count<br />

Sweep Count<br />

Features and Functions 3<br />

You can specify the number of triggers that will be accepted by the<br />

internal DMM before returning to the “idle” state. The trigger count applies<br />

to both scanning and stand- alone DMM measurements (with no scan list).<br />

Select a trigger count between 1 and 500,000 triggers, or continuous.<br />

You can store at least 500,000 readings in memory and all readings are<br />

automatically time stamped. If memory overflows, the new readings will<br />

overwrite the first (oldest) readings stored; the most recent readings are<br />

always preserved.<br />

You can specify a trigger count in conjunction with a sample count and<br />

a sweep count. The three parameters operate independent of one<br />

another, and the total number of readings returned will be the product<br />

of the three parameters.<br />

The CONFigure and MEASure? commands automatically set the scan<br />

trigger count to 1.<br />

The instrument sets the scan trigger count to 1 after a Factory Reset<br />

(*RST command). An Instrument Preset (SYSTem:PRESet command) or<br />

Card Reset (SYSTem:CPON command) does not change the setting.<br />

Front Panel Operation: Scan (Configure) > SCAN TRIGGER > COUNTED|INFINITE<br />

Remote Interface Operation: TRIGger:COUNt<br />

To configure a continuous scan, send TRIG:COUNT INFINITY.<br />

The sweep count sets the number of sweeps per trigger event during a<br />

scan (a sweep is one pass through the scan list). The front- panel sample<br />

annunciator (“ * ”) turns on during each measurement.<br />

Trigger<br />

Sweep count<br />

Sweep 1<br />

Sweep 2 Sweep n Trigger<br />

Sweep Count<br />

(1 to 500,000 sweeps)<br />

<strong>Agilent</strong> <strong>34980A</strong> Mainframe User’s Guide 117<br />

. . .<br />

t