3 Features and Functions<br />

Non-Sequential Scanning<br />

By default, the instrument scans the list of channels in ascending order<br />

from slot 1 through slot 8 (channels are reordered as needed). If your<br />

application requires non- ordered scanning of the channels in the present<br />

scan list, you can use the non- sequential scanning mode.<br />

This feature is available from the remote interface only.<br />

The scanning mode applies to the entire mainframe and cannot be<br />

selectively used on individual modules.<br />

When sequential scanning is enabled (default), the channels in the scan<br />

list are placed in ascending order from slot 1 through slot 8. Duplicate<br />

channels are not allowed. For example, (@2001,1003,1001,1003) will be<br />

interpreted as (@1001,1003,2001).<br />

When sequential scanning is disabled (OFF), the channels remain in the<br />

order presented in the scan list (see exception below). Multiple<br />

occurrences of the same channel are allowed. For example,<br />

(@2001,2001,2001) and (@3010,1003,1001,1005) are valid and the<br />

channels will be scanned in the order presented.<br />

When you specify a range of channels in the scan list, the channels are<br />

always sorted in ascending order, regardless of the scan order setting.<br />

Therefore, (@1009:1001) will always be interpreted as 1001, 1002, <br />

1003, etc.<br />

If you define a scan list with the sequential mode enabled and later<br />

disable the mode, the scan list will not be reordered; however, the scan<br />

list will be treated as a non- sequential list thereafter.<br />

If you have defined a scan list with the sequential mode disabled (OFF)<br />

and later enable the mode, the channels will be reordered.<br />

Non- sequential scan lists are not stored as part of the instrument state<br />

by the *SAV command; in this case, the ordered mode will be enabled<br />

and the scan list will be empty when the instrument state is restored<br />

(*RCL command).<br />

The scan order setting is stored in volatile memory and the ordered<br />

mode will be enabled when power is turned off or after a Factory Reset<br />

(*RST command).<br />

Remote Interface Operation: ROUTe:SCAN:ORDered {OFF|ON}<br />