3 Features and Functions<br />

Remote Interface Operation: The following command retrieves stored<br />

readings from memory (the readings are not erased).<br />

FETCh?<br />

Use the following commands to query the statistics on the readings stored<br />

in memory for a specific channel or from the internal DMM. These<br />

commands do not remove the data from memory.<br />

<br />

<br />

<br />

<br />

CALC:AVER:MIN? (@3005)<br />

CALC:AVER:MIN:TIME? (@3005)<br />

CALC:AVER:MAX? (@3005)<br />

CALC:AVER:MAX:TIME? (@3005)<br />

CALC:AVER:AVER? (@3005)<br />

CALC:AVER:COUNT? (@3005)<br />

CALC:AVER:PTPEAK? (@3005)<br />

The following command retrieves the last reading taken on channel 1 on<br />

the module in slot 3 during a scan.<br />

DATA:LAST? (@3001)<br />

The following command clears the contents of statistics memory for the<br />

selected channel.<br />

CALC:AVER:CLEAR (@3001)<br />

Use the following command to determine the total number of readings<br />

stored in memory (all channels) from the most recent scan.<br />

DATA:POINTS?<br />

The following command reads and clears the specified number of readings<br />

from memory. This allows you to continue a scan without losing data<br />

stored in memory (if memory becomes full, new readings will overwrite<br />

the first readings stored). The specified number of readings are cleared<br />

from memory, starting with the oldest reading.<br />

DATA:REMOVE? 12<br />

Minimum reading on channel<br />

Time minimum was logged<br />

<br />

Maximum reading on channel<br />

Time maximum was logged<br />

<br />

Average of all readings on channel<br />

<br />

Number of readings taken on channel<br />

<br />

Peak- to- peak (maximum–minimum)<br />