Quick Start AS <strong>3900</strong> 9<br />
Press keys: Description:<br />
[180] [Enter]<br />
[Enter]<br />
<br />
[Enter]<br />
<br />
[40] [Enter]<br />
[Enter]<br />
<br />
[Enter]<br />
<br />
[3] [Enter]<br />
[250] [Enter]<br />
Add another 180 µl Reagent–A to the<br />
Destination vial.<br />
Insert the third Mix step:<br />
Add 40 µl from the Sample vial to the<br />
Destination vial.<br />
Insert the fourth Mix step:<br />
Mix the contents of the Destination vial 3<br />
times with a volume of 250 µl<br />
Press [Escape] three times to go back to the Ready screen<br />
Programming the Series (See page 43 of this manual)<br />
Press keys: Description:<br />
[Series] [1] [Enter] Enter Series number 1.<br />
[1] [Enter] Enter Method number:1<br />
[1] [Enter] Enter First sample vial: 01<br />
[1] [Enter] Enter Last sample vial: 01<br />
[2] [Enter] Enter First destination vial: 02<br />
Press [Escape] (twice) to go back to the Ready screen<br />
Running the Series (See page 45 of this manual)<br />
Put a sample vial in position 1, an empty sealed destination vial in<br />
position 2 and a filled reagent vial in position 86. Be sure the reagent vial<br />
is filled correctly before starting a new Series. (If you can’t reach one of<br />
the tray positions, turn the tray manually).<br />
Press keys: Description:<br />
[Start/Stop]<br />
[1] [Enter] Start at Series number:1<br />
[1] [Enter] Stop after Series number: 1<br />
If you press AS <strong>3900</strong> will start searching for vial 86 and<br />
transport 180 µl to the destination vial twice, subsequently 40 µl of<br />
sample will be added and after mixing 3 times a 10 µl injection will be<br />
performed.