Injection Principle 19<br />

For the first injection, after a wash or<br />

after emptying of the buffer tubing.<br />

The syringe dispenser aspirates<br />

transport liquids from the transport vial<br />

to fill the sample line with transport<br />

liquid and remove wash solvents.<br />

Fig. 18 µl pick-up injection: filling sample lines with transport liquid.<br />

The needle moves from the transport<br />

vial to the sample vial. The injection<br />

valve is switched into the LOAD<br />

position.<br />

Fig. 19 µl pick-up injection: go to sample vial and switch injection valve to LOAD.<br />

Fig. 20 µl pick-up injection: filling the sample loop.<br />

The programmed injection volume is<br />

aspirated from the sample vial.<br />

The sample needle moves back to the<br />

transport vial.<br />

The sample is quantitatively<br />

transported into the loop, with<br />

transport liquid (mobile phase) from<br />

the transport vial.<br />

Fig. 21 µl pick-up injection: transporting sample into the sample loop.<br />

The injection valve switches into the<br />

INJECT position. The sample loop is<br />

now part of the HPLC mobile phase<br />

flow path: sample is transported to the<br />

column. The analysis starts.