Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007
Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007
Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
8 Quick Start AS <strong>3900</strong><br />
Press keys: Description:<br />
[1] [Enter] Enter Method number:1<br />
[1] [Enter] Enter First sample vial: 01<br />
[1] [Enter] Enter Last sample vial: 01<br />
Press [Escape] (twice) to go back to the Ready screen<br />
Running the Series (See page 45 of this manual)<br />
Put a sample vial in position 1 and a vial with transport solvent (mobile<br />
phase) in position 85.<br />
Be sure the transport vial is filled correctly before starting a new<br />
Series.<br />
Press keys: Description:<br />
[Start/Stop]<br />
[1] [Enter] Start at Series number:1<br />
[1] [Enter] Stop after Series number: 1<br />
If you press AS <strong>3900</strong> will start searching for the transport vial<br />
and perform 3 injections of 10 µl. After every injection the needle system<br />
will be washed with 1 syringe volume.<br />
EXAMPLE 3: Performing a 1:10 dilution followed by a 10 µl partial loop fill<br />
injection<br />
The mix method will perform the following:<br />
Transfer 360 µl from Reagent A to the destination vial, add 40 µl sample,<br />
mix 3 times with 250 µl and subsequently inject 10 µl.<br />
From the Ready screen:<br />
Programming the method (See page 38 of this manual)<br />
Injection Program:<br />
Press keys: Description:<br />
[Methods] [1] [Enter] Enter Method number 1.<br />
Selecting partial loop fill injection program<br />
[Enter]<br />
[30] [Enter] Flush volume: e.g. 30 µl<br />
[1] [Enter] Number of injections per vial: 1<br />
[10] [Enter] Injection volume: 10 µl<br />
[100] [Enter] Enter the analysis time e.g. 1 minute<br />
Press [Escape] once to go up one level.<br />
Wash Program:<br />
Press keys: Description:<br />
Selecting Wash program<br />
Disable the wash<br />
Press [Escape] once to go up one level.<br />
Mix Program: (See page 42 of this manual)<br />
Press keys: Description:<br />
Selecting Mix program<br />
Confirm the use of the Mix program<br />
<br />
[180] [Enter]<br />
[Enter]<br />
<br />
[Enter]<br />
Insert the first Mix step:<br />
Add 180 µl Reagent–A to the Destination<br />
vial.<br />
Insert the second Mix step: