8 Quick Start AS <strong>3900</strong><br />

Press keys: Description:<br />

[1] [Enter] Enter Method number:1<br />

[1] [Enter] Enter First sample vial: 01<br />

[1] [Enter] Enter Last sample vial: 01<br />

Press [Escape] (twice) to go back to the Ready screen<br />

Running the Series (See page 45 of this manual)<br />

Put a sample vial in position 1 and a vial with transport solvent (mobile<br />

phase) in position 85.<br />

Be sure the transport vial is filled correctly before starting a new<br />

Series.<br />

Press keys: Description:<br />

[Start/Stop]<br />

[1] [Enter] Start at Series number:1<br />

[1] [Enter] Stop after Series number: 1<br />

If you press AS <strong>3900</strong> will start searching for the transport vial<br />

and perform 3 injections of 10 µl. After every injection the needle system<br />

will be washed with 1 syringe volume.<br />

EXAMPLE 3: Performing a 1:10 dilution followed by a 10 µl partial loop fill<br />

injection<br />

The mix method will perform the following:<br />

Transfer 360 µl from Reagent A to the destination vial, add 40 µl sample,<br />

mix 3 times with 250 µl and subsequently inject 10 µl.<br />

From the Ready screen:<br />

Programming the method (See page 38 of this manual)<br />

Injection Program:<br />

Press keys: Description:<br />

[Methods] [1] [Enter] Enter Method number 1.<br />

Selecting partial loop fill injection program<br />

[Enter]<br />

[30] [Enter] Flush volume: e.g. 30 µl<br />

[1] [Enter] Number of injections per vial: 1<br />

[10] [Enter] Injection volume: 10 µl<br />

[100] [Enter] Enter the analysis time e.g. 1 minute<br />

Press [Escape] once to go up one level.<br />

Wash Program:<br />

Press keys: Description:<br />

Selecting Wash program<br />

Disable the wash<br />

Press [Escape] once to go up one level.<br />

Mix Program: (See page 42 of this manual)<br />

Press keys: Description:<br />

Selecting Mix program<br />

Confirm the use of the Mix program<br />

<br />

[180] [Enter]<br />

[Enter]<br />

<br />

[Enter]<br />

Insert the first Mix step:<br />

Add 180 µl Reagent–A to the Destination<br />

vial.<br />

Insert the second Mix step: