Troubleshooting 133<br />

Keine Injektion<br />

NO<br />

INJECTION<br />

Leakage inside<br />

Injection Valve<br />

Disconnect the needle tubing<br />

and the buffer tubing.<br />

Blockage<br />

in Flow<br />

Path<br />

Connect port 1 to an HPLC pump<br />

Block port 6.<br />

Start the pump at a low flow<br />

Abb. 49: Keine Injektion<br />

Start<br />

manual<br />

wash<br />

Disconnect<br />

Buffer tubing<br />

from valve<br />

Start<br />

manual<br />

wash<br />

Disconnect<br />

Buffer tubing<br />

from Syringe<br />

valve<br />

Start<br />

manual<br />

wash<br />

Disconnect<br />

needle from<br />

Valve<br />

Solvent<br />

flowing out of<br />

inj. port<br />

No<br />

Solvent<br />

out of open<br />

end<br />

No<br />

Solvent<br />

out of syringe<br />

valve<br />

Check for over tightened<br />

connections in the entire<br />

Flow Path &<br />

Check the syringe valve<br />

Observe<br />

ports 3 and 4<br />

for leakage<br />

No<br />

No<br />

Recheck with manual valve<br />

Yes<br />

Yes<br />

Yes<br />

Yes<br />

Check<br />

needle<br />

Check rotor<br />

seal<br />

Check<br />

buffer<br />

tubing<br />

Check rotor<br />

seal