24 Installation<br />

- With a syringe volume of wash solvent is aspirated<br />

from the wash solvent bottle and the wash solvent tubing is<br />

filled.<br />

- With the syringe contents will be dispensed to<br />

the syringe waste.<br />

• Repeat this action until the wash solvent tubing and the syringe are<br />

completely filled.<br />

• Press [Escape] to leave the maintenance screen.<br />

After the wash solvent tubing and the syringe are filled, press <br />

to perform a standard wash routine. All tubing connected to the syringe<br />

valve will be rinsed with wash solvent.<br />

Syringe<br />

The AS <strong>3900</strong> is factory equipped with a 250 µl syringe, but can also be<br />

equipped with a 1000 µl syringe, See page 55 how to replace the syringe.<br />

The AS <strong>3900</strong> will give the best results if all air is removed from the<br />

syringe. Use to remove air from the syringe, if this is not<br />

satisfactory, See page 23 for more tips on removing air.<br />

Preparation of vials<br />

Vial dimensions<br />

The AS <strong>3900</strong> is standard equipped with a tray which can contain 84<br />

standard sample and three large volume vials for reagents and transport<br />

liquid.<br />

Vials used should meet the following dimensions:<br />

Maximum vial height, including cap: 47 mm<br />

Minimum vial height, including cap: 32 mm<br />

Maximum vial diameter:<br />

Standard sample vial: 12 mm<br />

Reagent / transport liquid vials: 22 mm<br />

Large sample volume vials: 22 mm<br />

The following section contains examples of Chromacol vials witch can be<br />

used in the AS <strong>3900</strong>. See appendix A, List of accessories, for the<br />

corresponding article numbers for the vials, seals and caps.<br />

Chromacol standard sample vials<br />

The following Chromacol vials can be used in the AS <strong>3900</strong>. For each vial<br />

the Chromacol reference number is given.<br />

Fig. 26 Standard sample vials.