22 Installation<br />

Fluid connections<br />

RISK OF FIRE, REPLACE FUSES AS MARKED.<br />

When the voltage selection and fuses are correct for your power source,<br />

plug in the power cable.<br />

The AS <strong>3900</strong> is factory installed with a 250 µl syringe, a 100 µl loop, a<br />

500 µl buffer tubing and a stainless steel sample needle.<br />

A schematic presentation of all parts that have contact with any kind of<br />

fluid is shown below. This sticker can be found on the inside of the front<br />

cover.<br />

To get access to the fluidics, it is necessary to open the front cover of the<br />

AS <strong>3900</strong>, therefore remove the screw on the right-hand side of the AS<br />

<strong>3900</strong> front cover.<br />

See Table 1 for the dimension of the standard installed tubing.<br />

To obtain an optimal performance of the AS <strong>3900</strong>, a correct combination<br />

of syringe, loop and buffer tubing should be installed, see also section<br />

Needle and tubing on page 52.<br />

Fig. 25 Fluid connections of the AS <strong>3900</strong> (sticker).<br />

Table 1: Standard installed tubing of the AS <strong>3900</strong>.<br />

Tubing Material and dimensions Label<br />

Standard sample<br />

needle and tubing.<br />

Buffer tubing from<br />

high-pressure valve to<br />

syringe valve.<br />

Tubing syringe valve<br />

to wash solvent bottle.<br />

Tubing syringe valve<br />

to waste.<br />

SS tubing; 70 mm x 0.65 mm OD x<br />

0.25 mm ID<br />

Tefzel tubing; 155 mm x 1 /16” OD x<br />

0.25 mm ID<br />

PTFE tubing; 640 mm x 1 /16” OD x<br />

1.0 mm ID<br />

PTFE tubing; 300 mm x 1 /16” OD x<br />

1.0 mm ID<br />

PTFE tubing; 400 mm x 1 /16” OD x<br />

1.6 mm ID<br />

15 µl<br />

500 µl