70 APPENDIX C: Test Procedure<br />
After the dilution is completed, 10 µl is taken from both sample vials and<br />
injected three times each.<br />
Program the following methods and series on the autosampler:<br />
Method 1:<br />
Injection mode: Partial loop fill<br />
Flush volume: 30 µl<br />
No. of inj./vial: 3<br />
Injection volume: 10 µl<br />
Analysis time: 0:01:00<br />
Wash between: No<br />
Method 2:<br />
Injection mode: Partial loop fill<br />
Flush volume: 30 µl<br />
No. of inj./vial: 3<br />
Injection volume: 10 µl<br />
Analysis time: 0:01:00<br />
Wash between: No<br />
1) Add 90 µl Reag-A to DESTINATION<br />
2) Add 90 µl Reag-A to DESTINATION<br />
3) Add 40µl SAMPLE to DESTINATION<br />
4) Add 90 µl Reag-A to DESTINATION<br />
5) Add 90 µl Reag-A to DESTINATION<br />
MIX 3 times with 250 µl<br />
END OF MIX METHOD<br />
Series 1:<br />
Method: 1<br />
First vial: 01<br />
Last vial: 01<br />
Series 2:<br />
Method: 2<br />
First vial: 01<br />
Last vial: 01<br />
Destination vial: 02<br />
After measurement is completed, a chromatogram similar to Fig. 55<br />
should result.<br />
uAU<br />
700000<br />
600000<br />
500000<br />
400000<br />
300000<br />
200000<br />
100000<br />
0<br />
-100000<br />
-200000<br />
UV-Detektor[1]<br />
MIDAS_Mixing_Test<br />
Fig. 55: Mixture Test<br />
1 2 3 4 5 6 7 8 9 10 11<br />
Minutes<br />
700000<br />
600000<br />
500000<br />
400000<br />
300000<br />
200000<br />
100000<br />
0<br />
-100000<br />
-200000<br />
uAU