70 APPENDIX C: Test Procedure<br />

After the dilution is completed, 10 µl is taken from both sample vials and<br />

injected three times each.<br />

Program the following methods and series on the autosampler:<br />

Method 1:<br />

Injection mode: Partial loop fill<br />

Flush volume: 30 µl<br />

No. of inj./vial: 3<br />

Injection volume: 10 µl<br />

Analysis time: 0:01:00<br />

Wash between: No<br />

Method 2:<br />

Injection mode: Partial loop fill<br />

Flush volume: 30 µl<br />

No. of inj./vial: 3<br />

Injection volume: 10 µl<br />

Analysis time: 0:01:00<br />

Wash between: No<br />

1) Add 90 µl Reag-A to DESTINATION<br />

2) Add 90 µl Reag-A to DESTINATION<br />

3) Add 40µl SAMPLE to DESTINATION<br />

4) Add 90 µl Reag-A to DESTINATION<br />

5) Add 90 µl Reag-A to DESTINATION<br />

MIX 3 times with 250 µl<br />

END OF MIX METHOD<br />

Series 1:<br />

Method: 1<br />

First vial: 01<br />

Last vial: 01<br />

Series 2:<br />

Method: 2<br />

First vial: 01<br />

Last vial: 01<br />

Destination vial: 02<br />

After measurement is completed, a chromatogram similar to Fig. 55<br />

should result.<br />

uAU<br />

700000<br />

600000<br />

500000<br />

400000<br />

300000<br />

200000<br />

100000<br />

0<br />

-100000<br />

-200000<br />

UV-Detektor[1]<br />

MIDAS_Mixing_Test<br />

Fig. 55: Mixture Test<br />

1 2 3 4 5 6 7 8 9 10 11<br />

Minutes<br />

700000<br />

600000<br />

500000<br />

400000<br />

300000<br />

200000<br />

100000<br />

0<br />

-100000<br />

-200000<br />

uAU