32 Operation<br />

In the Ready screen the AS <strong>3900</strong> will display at the bottom line the soft<br />

functions which can be used from this screen. With the [Menu] key it is<br />

possible to toggle between the screens.<br />

At the right-hand side of the screen one or two temperatures are shown,<br />

the oven temperature and the tray temperature, when this option is<br />

installed.<br />

Direct Control functions:<br />

<br />

Activates the manual wash of the AS <strong>3900</strong>. All tubing, the needle and the<br />

syringe will be rinsed.<br />

<br />

The maintenance screens will provide three functions to control the<br />

needle arm, the syringe and the injection valve.<br />

<br />

This function is build-in to facilitate the replacement of the sample needle,<br />

air needle and valve. See also section Sample needle replacement on<br />

page 54, for detailed information.<br />

Press to move the needle arm horizontally above the inner<br />

sample ring. When the needle arm reaches the front position. It will return<br />

to its home position by pressing .<br />

<br />

This function facilitates the exchange of syringes and filling of the wash<br />

solvent tubing. See also section Syringe on page 55, for detailed<br />

information.<br />

Press to move the syringe to its end position, thereby<br />

aspirating wash solvent into the tubing and syringe. It stops from where it<br />

will return to its home position by pressing , thereby<br />

dispensing the wash solvent to waste.<br />

After changing a syringe all tubing can be rinsed using as<br />

described above.<br />

<br />

With this function the injection valve can be switched.<br />

Press to switch the valve from INJECT to LOAD.<br />

When the valve is in the LOAD position it can be switched back to<br />

INJECT by pressing .<br />

It is not possible to leave this function when the valve is in the LOAD<br />

position, first switch back by pressing and then press<br />

[Escape].<br />

To leave the maintenance screen and return to the Ready screen press<br />

[Escape].<br />

<br />

After pressing the AS <strong>3900</strong> enters the serial communication<br />

mode. From this mode the AS <strong>3900</strong> can be controlled from a personal<br />

computer using the RS232 interface and the <strong>Smartline</strong> application<br />

software. Refer to the <strong>Smartline</strong> manual for operating instructions. This<br />

manual is available on request as a separate document.