6 Quick Start AS <strong>3900</strong><br />
• After turning "ON" the power, the Ready screen appears displaying<br />
the firmware revision number.<br />
• Connect the drain tubing to the wash outlet on the lower front side<br />
of the instrument.<br />
• Fill the wash solvent bottle with distilled water/ 2-Propanol (80/20<br />
v/v%) or mobile phase. Only water/organic solvents should be<br />
used. Do not use any crystalline or buffer solutions; these may<br />
block the system and cause severe damages. To avoid air bubbles<br />
in the syringe degas the wash solvent.<br />
• Fill the wash solvent tubing, syringe and buffer tubing by washing<br />
the system 2 or 3 times. This is done by pressing the soft<br />
function key from the Ready screen. Check if there are some air<br />
bubbles trapped in the syringe, they can be removed by gently<br />
ticking to the syringe.<br />
• Connect your HPLC pump to port 1 of the injection valve and the<br />
column (or capillary) to port 6 of the injection valve. Check for<br />
leakage and let the system equilibrate for at least 5 minutes.<br />
System settings (See page 34 of this manual)<br />
Normally these settings are correctly installed, to check and set these<br />
(standard) settings, proceed as follows:<br />
Press keys: Description:<br />
[System] Enter the System Settings<br />
Enter the General Settings<br />
[0100] [Enter] Volume of installed loop: 100 µl.<br />
[015] [Enter] Volume of tubing needle ↔ valve:015 µl<br />
[Enter] Syringe volume: 250 µl<br />
[Enter] Set Syringe speed to Normal.<br />
[Enter] Skip missing vials.<br />
[Enter] Air segment: use air segment<br />
[Enter] Headspace pressure: On<br />
Press [Escape] twice to go back to the Ready screen<br />
Example 1: Performing a 10 µl partial loop fill injection from vial number 1.<br />
Performing injection routines is easy with AS <strong>3900</strong>. First you define a<br />
method such as the injection, mix or wash method. These methods are<br />
stored in battery back up. Subsequently a Run sequence or Series can<br />
be programmed in which those methods are linked to the Series of vials.<br />
AS <strong>3900</strong> will eventually execute those Series.<br />
Programming the method (See page 38 of this manual)<br />
Injection Program:<br />
Press keys: Description:<br />
[Methods] [1] [Enter] Enter Method number 1.<br />
Selecting partial loop fill injection<br />
[Enter]<br />
program<br />
[30] [Enter] Flush volume: e.g. 30 µl<br />
[1] [Enter] Number of injections per vial: 1<br />
[10] [Enter] Injection volume: 10 µl<br />
[100] [Enter] Enter the analysis time e.g. 1 minute<br />
Press [Escape] once to go up one level.<br />
Wash Program:<br />
Press keys: Description:<br />
Selecting wash program<br />
Disable the wash<br />
Press [Escape] three times to go back to the Ready screen.