12.02.2013 Views

Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007

Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007

Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Installation 29<br />

Multi Link Connector<br />

The MultiLink connector is used to make the AS <strong>3900</strong> communicate with<br />

software packages like ChromGate ® and EuroChrom ® .<br />

S1<br />

ON DIP<br />

S2<br />

ON DIP<br />

OUT IN<br />

Fig. 34 MultiLink board<br />

Set the dipswitch S1 and S2 as follows to let the communication work<br />

probably:<br />

S1 1 2 3 4<br />

ON<br />

OFF<br />

S2 1 2 3 4<br />

Port Description<br />

S1: OUT to next device<br />

S2: IN to PC or previous device<br />

Connections for 9 pin connector:<br />

ON<br />

OFF<br />

pin 2 TD Transmitted data to the computer.<br />

pin 3 RD Received data from the computer.<br />

pin 5 SG<br />

Signal ground (Also indicated as GND in<br />

some devices)