14 Injection Principle<br />
1. P2- I/O connector; Inputs, see page 27.<br />
2. P1- I/O connector; Relays outputs.<br />
3. Mains switch.<br />
4. Mains input.<br />
5. Fuses and voltage selector.<br />
6. CE-mark.<br />
7. RS-232 communication interface connector.<br />
8. Fan, only when tray-cooling option is installed.<br />
Fig. 4 Rear side of the autosampler AS <strong>3900</strong>.<br />
Injection Principle<br />
A unique sequence of valve switching and syringe dispenser controlled<br />
sample withdrawal enables the AS <strong>3900</strong> to inject sample volumes in a<br />
range from a few microliters to milliliters in a simple and reproducible<br />
manner.<br />
Therefore three different methods for injection can be selected:<br />
• Full loop: The sample loop is completely (quantitatively)<br />
filled with sample resulting in extremely good<br />
reproducibility.<br />
• Partial loop fill: The sample loop is partially filled with sample;<br />
low sample loss and programmable injection<br />
volumes.<br />
• µl pick-up: After aspiration from the vial the sample volume<br />
is transported into the loop with transport liquid<br />
(mobile phase) from another vial; no sample loss.<br />