14 Injection Principle<br />

1. P2- I/O connector; Inputs, see page 27.<br />

2. P1- I/O connector; Relays outputs.<br />

3. Mains switch.<br />

4. Mains input.<br />

5. Fuses and voltage selector.<br />

6. CE-mark.<br />

7. RS-232 communication interface connector.<br />

8. Fan, only when tray-cooling option is installed.<br />

Fig. 4 Rear side of the autosampler AS <strong>3900</strong>.<br />

Injection Principle<br />

A unique sequence of valve switching and syringe dispenser controlled<br />

sample withdrawal enables the AS <strong>3900</strong> to inject sample volumes in a<br />

range from a few microliters to milliliters in a simple and reproducible<br />

manner.<br />

Therefore three different methods for injection can be selected:<br />

• Full loop: The sample loop is completely (quantitatively)<br />

filled with sample resulting in extremely good<br />

reproducibility.<br />

• Partial loop fill: The sample loop is partially filled with sample;<br />

low sample loss and programmable injection<br />

volumes.<br />

• µl pick-up: After aspiration from the vial the sample volume<br />

is transported into the loop with transport liquid<br />

(mobile phase) from another vial; no sample loss.<br />

S1<br />

ON DIP<br />

S2<br />

ON DIP<br />

OUT IN