36 Operation<br />

Table 4: Syringe speed.<br />

250 µl syringe 1000 µl syringe<br />

Syringe speed: Aspirate Dispense Aspirate Dispense<br />

LOW 0.31 ml/min 3.4 ml/min 1.25 ml/min 5.5 ml/min<br />

NORMAL 0.62 ml/min 3.4 ml/min 2.50 ml/min 5.5 ml/min<br />

HIGH 0.94 ml/min 3.4 ml/min 3.75 ml/min 5.5 ml/min<br />

Wash syringe<br />

speed:<br />

Method System Settings<br />

Press :<br />

METHOD<br />

SYSTEM<br />

SETTINGS<br />

Use calibration<br />

vials:<br />

Tray Settings<br />

Press :<br />

TRAY<br />

SETTINGS<br />

Range<br />

3.4 ml/min<br />

<br />

<br />

Comment<br />

Range Comment<br />

13.7 ml/min<br />

Determine if calibration vials<br />

can be used in Series.<br />

Tray type: Standard tray with 84 vials, 1.5<br />

ml and 3 vials of 10 ml for<br />

special use.<br />

Vials 85-87, will be used for<br />

transport solvent for the µl pickup<br />

injections.<br />

When the use of mix item is<br />

enabled in the System Setting<br />

the three 10 ml vials will be used<br />

as follows:<br />

vial 85:<br />

vial 86:<br />

vial 87:<br />

transport solvent<br />

Reagent A<br />

Reagent B<br />

Tray type: Tray with 96 vials of 1.5 ml.<br />

Tray type: Tray with 24 vials of 10 ml.<br />

Vial type: <br />

Standard vials are used.<br />

(only when Only 2.5 ml vials are used.<br />

or<br />

(Chromacol 2.5-CV, 12 X 40<br />

is<br />

selected)<br />

mm)<br />

Only with the standard tray it is possible to perform a mix<br />

and to use the µl pick-up injection mode.<br />

With the other two trays both, the mix and the µl pick-up injection<br />

mode will automatically be disabled.<br />

If 2.5 ml vials are used be sure to fill the vials completely before<br />

starting a RUN. AS <strong>3900</strong>.