38 Operation<br />
For information how to install the Prep option to a standard AS <strong>3900</strong> see<br />
appendix E.<br />
Injection volumes Prep AS<br />
Ensure that the programmed injection volume does not exceed the<br />
50% of the loop volume, otherwise the reproducibility and accuracy<br />
of the injection can not be guaranteed.<br />
Method programming<br />
The AS <strong>3900</strong> offers nine programmable methods.<br />
Depending on your needs a method may comprise:<br />
• Injection program Containing information on the type of the<br />
injection with the flush volume, injection<br />
volume, number of injections per vial and<br />
the analysis time.<br />
• Wash program Containing information on the wash of the<br />
AS <strong>3900</strong>.<br />
• Mix program Containing information on the sample<br />
handling of the sample prior to the injection.<br />
• Timed events program Containing information on the auxiliaries and<br />
the temperature of the column oven during<br />
the analysis of the sample.<br />
A method is any combination of the above and is saved as combination<br />
under its method number.<br />
To enter the method programming mode press [Method].<br />
After entering the method number the AS <strong>3900</strong> will return with the main<br />
screen of method programming:<br />
PROGRAMMING METHOD NUMBER: 1 (I W M T)<br />
SELECT SECTION TO PROGRAM<br />
<br />
From this screen the different sections can be programmed after pressing<br />
the corresponding soft function key.<br />
Existing programs in this method are displayed between brackets on the<br />
first line; I for the injection program, W for the wash program, M for the<br />
mix program and T for the timed events program.<br />
After programming the method press [Escape] to leave the method<br />
programming.<br />
The methods are stored in battery backup memory.<br />
Injection program<br />
After pressing the AS <strong>3900</strong> returns with the main screen<br />
of injection programming.<br />
METHOD NUMBER: 1 SECTION: INJECTION<br />
INJECTION MODE: PARTIAL LOOPFILL<br />