12.02.2013 Views

Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007

Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007

Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

18 Injection Principle<br />

Fig. 15 Partial loop fill injection: injecting sample.<br />

The injection valve switches into the<br />

INJECT position. The sample loop is<br />

now part of the HPLC mobile phase<br />

flow path: sample is transported to the<br />

column. The analysis starts.<br />

The next injection sequence will start with a flush of 50% of the<br />

programmed flush volume, in case an injection from the same vial and no<br />

wash routine is programmed. Otherwise it will start with a flush of the<br />

programmed flush volume. If the withdrawal of sample for the next<br />

injection will exceed the total volume of the sample buffer tubing, the<br />

buffer tubing is rinsed before the next injection. The next injection will<br />

start with the programmed flush.<br />

Air segment<br />

An air segment can be used to reduce the amount of flush volume. This<br />

air segment is at the front of the flush volume and will not be injected.<br />

The air segment can be selected in system settings (ON/OFF, See page<br />

34).<br />

With a standard needle the flush volumes must be minimal 30 µl for<br />

injections with air segment and 35 µl for injections without air segment. If<br />

the samples are highly viscous it may be necessary to program larger<br />

flush volumes and reduce the syringe speed for better performance.<br />

Fig. 16 Situation after filling the loop in case of Partial loop fill injections with (A)<br />

or without air segment (B)<br />

µl Pick-up injections<br />

The switching sequence for a µl pick-up injection is schematically shown<br />

in the following figures:<br />

The Initial situation: the injection valve<br />

is in the INJECT position. The sample<br />

needle has entered the vial of<br />

transport liquid (mobile phase, this to<br />

avoid disturbance of the<br />

chromatogram with an additional peak<br />

of the transport solvent).<br />

Fig. 17 µl pick-up injection: initial situation.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!