Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007
Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007
Smartline Autosampler 3900 Manual / Handbuch V 1508 03/2007
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
18 Injection Principle<br />
Fig. 15 Partial loop fill injection: injecting sample.<br />
The injection valve switches into the<br />
INJECT position. The sample loop is<br />
now part of the HPLC mobile phase<br />
flow path: sample is transported to the<br />
column. The analysis starts.<br />
The next injection sequence will start with a flush of 50% of the<br />
programmed flush volume, in case an injection from the same vial and no<br />
wash routine is programmed. Otherwise it will start with a flush of the<br />
programmed flush volume. If the withdrawal of sample for the next<br />
injection will exceed the total volume of the sample buffer tubing, the<br />
buffer tubing is rinsed before the next injection. The next injection will<br />
start with the programmed flush.<br />
Air segment<br />
An air segment can be used to reduce the amount of flush volume. This<br />
air segment is at the front of the flush volume and will not be injected.<br />
The air segment can be selected in system settings (ON/OFF, See page<br />
34).<br />
With a standard needle the flush volumes must be minimal 30 µl for<br />
injections with air segment and 35 µl for injections without air segment. If<br />
the samples are highly viscous it may be necessary to program larger<br />
flush volumes and reduce the syringe speed for better performance.<br />
Fig. 16 Situation after filling the loop in case of Partial loop fill injections with (A)<br />
or without air segment (B)<br />
µl Pick-up injections<br />
The switching sequence for a µl pick-up injection is schematically shown<br />
in the following figures:<br />
The Initial situation: the injection valve<br />
is in the INJECT position. The sample<br />
needle has entered the vial of<br />
transport liquid (mobile phase, this to<br />
avoid disturbance of the<br />
chromatogram with an additional peak<br />
of the transport solvent).<br />
Fig. 17 µl pick-up injection: initial situation.