APPENDIX D: AS <strong>3900</strong>, Preparative 73<br />

APPENDIX D: AS <strong>3900</strong>, Preparative<br />

By just choosing PREP in your system settings, you can use the AS <strong>3900</strong><br />

to inject all of your samples into a Preparative LC system or in other<br />

areas where large injection volumes are required. Also in combination<br />

with the PROSPEKT On-line Solid Phase Extraction System, the<br />

AS <strong>3900</strong> can be your ideal autosampler.<br />

The combination of large sample vials (10 ml) a large sample volume<br />

needle and a 2.5 ml syringe enables you to inject large volumes very<br />

reproducible with high speeds and only 45 µl of sample loss. The<br />

installed large bore valve (0.75 mm) with 10 ml sample loop enables you<br />

to inject from microliters to milliliters with the same autosampler. Flow<br />

rates up to 200 ml per minute are no problem for the autosampler<br />

AS <strong>3900</strong>, used in Prep mode.<br />

If the Prep option is factory installed the part on installation can be<br />

skipped.<br />

If the Prep option is bought as a kit, carry out the installation instructions.<br />

Specifications AS <strong>3900</strong>, prep<br />

The Specifications of the AS <strong>3900</strong>, prep differ in some ways from the<br />

standard autosampler AS <strong>3900</strong>. The different specifications are listed<br />

below.<br />

SAMPLING<br />

Sample capacity<br />

LSV tray: 24 vials of 10 ml (LSV)<br />

Loop volume Not programmable, injection volume<br />

determines the aspirated sample<br />

volume<br />

Dispenser syringe 2500 µl syringe<br />

ANALYTICAL PERFORMANCE<br />

Reproducibility RSD 10 µl<br />

up to 50% of the installed sample<br />

loop.<br />

Memory effect < 0.1% with programmable needle<br />

wash<br />

PROGRAMMING<br />

Methods 9 programmable methods<br />

Injection method Partial loop fill only<br />

Injection volume 1 µl – 19,999 µl, with 1 µl increment<br />

Mix Not available<br />

OPTIONS<br />

Bio-compatible valve Not available