APPENDIX D: AS <strong>3900</strong>, Preparative 73<br />
APPENDIX D: AS <strong>3900</strong>, Preparative<br />
By just choosing PREP in your system settings, you can use the AS <strong>3900</strong><br />
to inject all of your samples into a Preparative LC system or in other<br />
areas where large injection volumes are required. Also in combination<br />
with the PROSPEKT On-line Solid Phase Extraction System, the<br />
AS <strong>3900</strong> can be your ideal autosampler.<br />
The combination of large sample vials (10 ml) a large sample volume<br />
needle and a 2.5 ml syringe enables you to inject large volumes very<br />
reproducible with high speeds and only 45 µl of sample loss. The<br />
installed large bore valve (0.75 mm) with 10 ml sample loop enables you<br />
to inject from microliters to milliliters with the same autosampler. Flow<br />
rates up to 200 ml per minute are no problem for the autosampler<br />
AS <strong>3900</strong>, used in Prep mode.<br />
If the Prep option is factory installed the part on installation can be<br />
skipped.<br />
If the Prep option is bought as a kit, carry out the installation instructions.<br />
Specifications AS <strong>3900</strong>, prep<br />
The Specifications of the AS <strong>3900</strong>, prep differ in some ways from the<br />
standard autosampler AS <strong>3900</strong>. The different specifications are listed<br />
below.<br />
SAMPLING<br />
Sample capacity<br />
LSV tray: 24 vials of 10 ml (LSV)<br />
Loop volume Not programmable, injection volume<br />
determines the aspirated sample<br />
volume<br />
Dispenser syringe 2500 µl syringe<br />
ANALYTICAL PERFORMANCE<br />
Reproducibility RSD 10 µl<br />
up to 50% of the installed sample<br />
loop.<br />
Memory effect < 0.1% with programmable needle<br />
wash<br />
PROGRAMMING<br />
Methods 9 programmable methods<br />
Injection method Partial loop fill only<br />
Injection volume 1 µl – 19,999 µl, with 1 µl increment<br />
Mix Not available<br />
OPTIONS<br />
Bio-compatible valve Not available