Installation 25<br />
Fig. 27 Conical vials with support.<br />
Fig. 28 Plastic vials.<br />
To reduce the volume of standard vials use a Chromacol insert (02–MTV,<br />
02–MTVWG or <strong>03</strong>–MTV) in combination with the appropriate vial and<br />
support sleeve or spring.<br />
Fig. 29 Standard vial with insert.<br />
Chromacol 2.5-CV vials<br />
It is possible to use the Chromacol 2.5-CV vials in the AS <strong>3900</strong>.<br />
When used, this vial type must be programmed in the System Settings,<br />
see page 34.<br />
AS <strong>3900</strong> uses two aspiration heights to aspirate liquid from the 2.5-<br />
CV vials. Therefore the vials need to be full, when AS <strong>3900</strong> is<br />
started.<br />
When a vial is not full the AS <strong>3900</strong> might aspirate air!<br />
Chromacol reagent and transport liquid vial<br />
The following Chromacol vial can be used for reagents and transport<br />
liquid; this vial can also be used in the optional LSV tray.<br />
Fig. 30 Reagent and transport liquid vial.