60 Troubleshooting<br />
Fig. 50:<br />
Bad reproducibility<br />
BAD<br />
REPRODUCIBILITY<br />
Air in Flow Path Start manual wash<br />
Air in Syringe<br />
Leaking Syringe<br />
Leaking Syringe<br />
Valve<br />
Rotor seal worn out<br />
Dead volumes in<br />
tubing connections<br />
Remove Syringe and fill<br />
manually with wash solvent<br />
Replace plunger tip or<br />
Syringe<br />
Replace Valve<br />
Replace seal<br />
Check Stator<br />
Redo connections with new<br />
Ferrules and nuts