34 Operation<br />
Turn the tray cooling on by pressing and program the new set<br />
point. After pressing [Enter] the new tray cooling set point will be<br />
activated and the AS <strong>3900</strong> returns to the first temperature-programming<br />
screen.<br />
Temperature range tray cooling: 4°C - 15°C<br />
Maximum cool capacity: 20°C below ambient temperature<br />
After all values are programmed press [Escape] to return to the Ready<br />
screen.<br />
<br />
After pressing the AS <strong>3900</strong> offers three Utility soft functions:<br />
Copy a method.<br />
Erase a method.<br />
AS <strong>3900</strong> logbook, containing information on<br />
the number of switching of the injection valve<br />
and the syringe valve and on the number of<br />
movements of the syringe dispenser.<br />
<br />
The service mode of the AS <strong>3900</strong> can be entered after pressing<br />
and entering the service entry-code. This code is only<br />
known to authorized service personnel.<br />
System Settings<br />
After pressing [System] the AS <strong>3900</strong> enters the System Settings, where<br />
all system variables of the AS <strong>3900</strong> can be set.<br />
[Menu]<br />
SYSTEM SETTINGS [MENU]<br />
SELECT SETTINGS TO DEFINE<br />
<br />
SYSTEM SETTINGS [MENU]<br />
SELECT SETTINGS TO DEFINE<br />
TRAY:10°C<br />
<br />
The System Settings are divided in five sub-sections: to enter a subsection,<br />
press the corresponding soft function key. To step through a subsection<br />
press the [Enter] key.<br />
After all values are defined correctly within a sub-section, return the menu<br />
screen of the System Settings by pressing [Escape]. Return to the<br />
Ready screen by pressing [Escape] one more time. The values are<br />
stored in battery back up memory.<br />
The following sections show the programmable system settings, with their<br />
ranges.