44 Operation<br />
Should a method consist of only the wash and/or a timed events<br />
program, no more questions will be asked.<br />
When the use of calibration vials is enabled in the System Settings (See<br />
page 34) and the method does not contain a mix, the AS <strong>3900</strong> will ask<br />
the following questions:<br />
Use calibration vials (see Fig. 36) <br />
: • First calibration vial:<br />
• Last calibration vial:<br />
• No of samples between calibrations<br />
The range of sample vials is:<br />
• Last sample vial<br />
• Last sample vial<br />
When the method contains a mix item in which destination vials are used,<br />
the AS <strong>3900</strong> will ask for the first position of the destination vials. The last<br />
destination vial is calculated based on the sample vial range:<br />
• First destination vial:<br />
Fig. 36: Run sequence with 3 calibration vials between every 5 vials (First<br />
calibration vial: 1, last calibration vial: 3 and number of samples between<br />
calibrations: 5).<br />
Sample vials and destination vials must be placed sequentially.<br />
Sample vial and destination vial ranges must not overlap.<br />
After all parameters are programmed return to the Series programming<br />
screen by pressing [Escape], where an other Series can be<br />
programmed. If all Series are programmed, return to the Ready screen<br />
by pressing [Escape].<br />
Series are not stored in battery backup memory.<br />
After switching ON the Series are empty (method: ).