44 Operation<br />

Should a method consist of only the wash and/or a timed events<br />

program, no more questions will be asked.<br />

When the use of calibration vials is enabled in the System Settings (See<br />

page 34) and the method does not contain a mix, the AS <strong>3900</strong> will ask<br />

the following questions:<br />

Use calibration vials (see Fig. 36) <br />

: • First calibration vial:<br />

• Last calibration vial:<br />

• No of samples between calibrations<br />

The range of sample vials is:<br />

• Last sample vial<br />

• Last sample vial<br />

When the method contains a mix item in which destination vials are used,<br />

the AS <strong>3900</strong> will ask for the first position of the destination vials. The last<br />

destination vial is calculated based on the sample vial range:<br />

• First destination vial:<br />

Fig. 36: Run sequence with 3 calibration vials between every 5 vials (First<br />

calibration vial: 1, last calibration vial: 3 and number of samples between<br />

calibrations: 5).<br />

Sample vials and destination vials must be placed sequentially.<br />

Sample vial and destination vial ranges must not overlap.<br />

After all parameters are programmed return to the Series programming<br />

screen by pressing [Escape], where an other Series can be<br />

programmed. If all Series are programmed, return to the Ready screen<br />

by pressing [Escape].<br />

Series are not stored in battery backup memory.<br />

After switching ON the Series are empty (method: ).