Chapter 6. MVS-<strong>STX</strong> Inputs <strong>and</strong> Outputs Menu<br />

11. The display looks like this:<br />

Addr XX CHAN# YY<br />

420mA*<br />

020mA<br />

F1 F2 F3<br />

(Asterisk indicates the current selection.)<br />

12. Press the F1 Key for the current output<br />

range to be 4 to 20 mA or the F3 Key for<br />

0 to 20 mA. The display acknowledges<br />

the selection <strong>and</strong> returns to:<br />

CURRENT OUTPUT<br />

Set Mode Rprt<br />

F1 F2 F3<br />

13. Press the F1 Key to access the Set<br />

Menu. If you added one output, proceed<br />

to Step 14. If you added two outputs, the<br />

display looks like this:<br />

SELECT OUTPUT<br />

X/20#1<br />

Y/20#2<br />

F1 F2 F3<br />

A ‘4’ or ‘0’ will be in place of the ‘X’ <strong>and</strong><br />

‘Y.’ Press the F1 Key to set up output #1<br />

or the F3 Key to set up output #2.<br />

14. The display looks like this:<br />

SET:<br />

XmA<br />

A ‘4’ or a ‘0’ will be in place of the ‘X.’<br />

15. Press the F1 Key to access 4/0mA. The<br />

display looks like this:<br />

X/20#Y LO VALUE<br />

A ‘4’ or a ‘0’ will be in place of the ‘X.’<br />

(Units are consistent with Units Menu.)<br />

16. Use the keypad or Arrow Keys to input a<br />

value for the weight where you want the<br />

current transmitter to output 4 mA (0 mA<br />

if applicable). Press the Enter Key.<br />

The display acknowledges the entry <strong>and</strong><br />

returns to:<br />

SET:<br />

XmA<br />

20mA<br />

F1 F2 F3<br />

> 0 lbs<br />

F1 F2 F3<br />

20mA<br />

F1 F2 F3<br />

17. Press the F3 Key to access 20mA. The<br />

display looks like this:<br />

X/20#Y HI VALUE<br />

18. Use the keypad or Arrow Keys to input a<br />

value for the weight where you want the<br />

current transmitter to output 20 mA.<br />

Press the Enter Key. The display acknowledges<br />

the entry <strong>and</strong> returns to:<br />

SET:<br />

XmA<br />

F1 F2 F3<br />

19. Press the Menu Key to access the<br />

menu’s second page. The display shows:<br />

SET:<br />

Net/Grs<br />

22. Press the F3 Key to select Fsafe. The<br />

display shows:<br />

FAIL-SAFE MODE?<br />

Lo* Hi NC<br />

F1 F2 F3<br />

(Asterisk indicates the current selection.)<br />

23. Press the F1, F2, or F3 Key to select the<br />

desired fail-safe condition. The display<br />

acknowledges the selection <strong>and</strong><br />

returns to:<br />

SET:<br />

Net/Grs<br />

FSafe<br />

F1 F2 F3<br />

20. Press the F1 Key to access Net/Grs.<br />

The display shows:<br />

SELECT OPERATION<br />

Net<br />

*Gross<br />

F1 F2 F3<br />

(Asterisk indicates the current selection.)<br />

21. Press the F1 Key to select Net or the F3<br />

Key to select Gross. The display acknowledges<br />

the selection <strong>and</strong> returns to:<br />

SET:<br />

Net/Grs<br />

20mA<br />

FSafe<br />

F1 F2 F3<br />

FSafe<br />

F1 F2 F3<br />

24. Press the Esc Key to return to the<br />

Current Output Menu <strong>and</strong> repeat Steps<br />

10 through 23 to set up additional outputs<br />

if desired.<br />

25. Press the Esc Key to scroll up the menu<br />

tree or press the Auto/Man Key to return<br />

to channel monitoring.<br />

> 9999 lbs<br />

F1 F2 F3<br />

A ‘4’ or a ‘0’ will be in place of the ‘X.’<br />

(Units are consistent with Units Menu.)<br />

6-8