Appendix B. Serial Comm<strong>and</strong>s<br />
Note<br />
The remaining comm<strong>and</strong>s deal with performing an Auto Calibration. Refer to the Auto section<br />
in Chapter 7, MVS-<strong>STX</strong> Calibration Menu, for a description of the calibration functions<br />
<strong>and</strong> methods.<br />
‘L’ Comm<strong>and</strong> — Auto Lo Span Calibration<br />
This comm<strong>and</strong> is sent by the master to request the <strong>STX</strong> to set Lo Span for the channel. For this<br />
example, the master sets the Lo Span to 25,000 lbs for an <strong>STX</strong> at address 01. The <strong>STX</strong> returns a<br />
status code of 0 (calibration successful) <strong>and</strong> a checksum of 30.<br />
Request from Master<br />
Response from <strong>STX</strong><br />
>aaLdddddddssr Adssr Format<br />
>01L002500004r A030r Example<br />
Note: In response from <strong>STX</strong>:<br />
d = 0 = calibration successful<br />
d = 1 = “Ambiguous” warning. Hi span counts are greater than lo span counts.<br />
d = 2 = “Move more material” warning. Change in weight caused a change of less than<br />
20,000 counts.<br />
‘H’ Comm<strong>and</strong> — Auto Hi Span Calibration<br />
This comm<strong>and</strong> is sent by the master to request the <strong>STX</strong> to set Hi Span for the channel. For this<br />
example, the master sets the Hi Span to 50,000 lbs for an <strong>STX</strong> at address 01. The <strong>STX</strong> returns a<br />
status code of 0 (calibration successful) <strong>and</strong> a checksum of 30.<br />
Request from Master<br />
Response from <strong>STX</strong><br />
>aaHdddddddssr Adssr Format<br />
>01H0050000FEr A030r Example<br />
Note: In response from <strong>STX</strong>:<br />
d = 0 = calibration successful<br />
d = 1 = “Ambiguous” warning. Hi Span counts are greater than Lo Span counts.<br />
d = 2 = “Move more material” warning. Change in weight caused a change of less than<br />
20,000 counts.<br />
‘Z’ Comm<strong>and</strong> — Auto Zero Calibration<br />
This comm<strong>and</strong> is sent by the master to request the <strong>STX</strong> to set Zero for the channel. For this<br />
example, the master sets the Zero Calibration point to 0 lbs for an <strong>STX</strong> at address 01. The <strong>STX</strong><br />
returns a status code of 0 (calibration successful) <strong>and</strong> a checksum of 30.<br />
Request from Master<br />
Response from <strong>STX</strong><br />
>aaZdddddddssr Adssr Format<br />
>01Z00000000Br A030r Example<br />
Note: In response from <strong>STX</strong>: d = 0 = calibration successful<br />