158 - BOOK 1 - WELLCAD BASICS<br />

www.alt.lu<br />

� Select Process Spectral Gamma Data from Process > Common ><br />

Spectral Gamma.<br />

� The process tries to reload the last used model file. If it cannot be found<br />

the user will be informed.<br />

� The Spectral Gamma Model Processing dialog box will open.<br />

Log, containing the data<br />

to process.<br />

Information about the<br />

type of model and<br />

remarks on the<br />

calibration data.<br />

SGM model file loaded into Spectral Gamma model Processing dialog<br />

From the drop down list in the Input section, select the log that contains the data<br />

you would like to process and use the Load Model button to find another model<br />

file.<br />

Currently two types of models, both saved in SGM files, can be processed. Models<br />

of type Output counts per energy window contain the low and high limits of<br />

energy windows only and will extract the total counts per energy window and output<br />

the result in separate Well Logs. You will also get the standard deviation for each<br />

window.<br />

If a model of type Spectrum stripping using energy window is contained in the<br />

SGM file a calibration of the count rates per energy window will be applied<br />

according to the calibration coefficients stored in the model file. One Well Log will<br />

be created per calibration model used for the master calibration. The values are<br />

output in calibrated units (e.g. pCi/g, Bq/g or ppm). If the model was setup to contain<br />

conversion factors from calibration units into other units (e.g. conversion from<br />

pCi/g to ppm or %) new Well Logs containing the converted values will be generated<br />

as well. E.g. a model file created at a KUT calibration pit with three different<br />

enriched zones of KUT concentrations/decay rates provided in pCi/g will output<br />

three Well Logs containing K, U and Th concentrations in pCi/g determined from