6. Defining the mounting positions<br />
6.1 General requirements<br />
All <strong>de</strong>vices for the nurse call<br />
system<br />
Devices with displays<br />
Call switches, Call <strong>de</strong>vices<br />
Presence buttons<br />
Optical signalling <strong>de</strong>vices<br />
Signal lamps, Corridor displays<br />
- Consi<strong>de</strong>ring the different users, the process of raising a call<br />
must be simple and unambiguous, and a mix-up with other system<br />
elements must be virtually impossible.<br />
- Damage or <strong>de</strong>struction from external effects which may occur<br />
during normal use (e.g. when moving beds) must be virtually<br />
impossible.<br />
- Must not be installed in rooms with harmfull influences to telecommunications<br />
equipment.<br />
- Unobstructed view must be ensured (ComTerminal, Room‐<br />
Terminal, ComStation CT , ComStation T ).<br />
- Not subject to direct sun radiation.<br />
- Easy access and reach must be ensured.<br />
- Near the door.<br />
- Visual recognition must not be obscured by other stray light.<br />
- Called or paged persons shall be directed via the shortest<br />
distance to the call location.<br />
- Shall be easily allocated to the respective rooms.<br />
Room lamps<br />
In WCs and wet rooms - Comply with specifications acc. to DIN VDE 0100!<br />
Power supply units,<br />
OSYlink modules,<br />
OSYnet-SafetyController,<br />
OSYnet‐Y‐RepeaterOpto,<br />
OSYnet‐Gateway<br />
Distributor cabinets for both<br />
nurse call system and mains<br />
power supply<br />
- Use in dry rooms only; if possible within distribution cabinets<br />
(not insi<strong>de</strong> of patient or resi<strong>de</strong>nt rooms).<br />
- Easy access at any time (access door at least 60 cm wi<strong>de</strong>).<br />
- Heat transfer must not be blocked off.<br />
- When installed in switch cabinets, forced-air cooling may be<br />
required.<br />
- Consi<strong>de</strong>ring the effective voltage drop, the power supply unit<br />
should be located close to the biggest users.<br />
- Fire hazard! Danger of fire in case of short-circuit. Therefore,<br />
ensure sufficient ventilation in distribution cabinets.<br />
- After removing the outer cover, the high-voltage part must still<br />
be protected.<br />
- Connection terminals for the nurse call system and those for<br />
the high-voltage installations must be clearly separated and<br />
<strong>de</strong>signated as such, e.g. by way of <strong>de</strong>sign, shape, colour. Inscriptions<br />
alone are not sufficient to preclu<strong>de</strong> misuse.<br />
6.1.1 Mounting heights<br />
Prescribed mounting height acc. to DIN VDE 0834-1:2000:<br />
Mounting height above floor level:<br />
Device with control elements (with and without indicating lamps) 700 - 1500 mm<br />
Device with control elements with text display<br />
1500 - 1700 mm<br />
Device in installation units<br />
1600 - 1800 mm<br />
Signal lamps and large letter display<br />
1500 - 2200 mm<br />
Distribution boxes<br />
700 - 2200 mm<br />
<strong>Flamenco</strong> - <strong>Technical</strong> <strong>Manual</strong>, 04/13 (Rev. <strong>6.0</strong>), 00 8812 60<br />