12. Functional checks<br />
12.3 Checking the consoles<br />
Perform this test for all installed consoles: Ward consoles, mobile consoles,<br />
central consoles. For information on the correct operation and use of the consoles<br />
refer to the relevant operating instructions.<br />
Two persons are required for executing these checks. One person will handle<br />
the console, and the other person will proceed from one room to another.<br />
Check for each room:<br />
Is the room number correctly displayed?<br />
Do all types of calls arrive at the console and are they signalled in the appropriate<br />
manner (e.g. normal call, WC call, diagnostic call, emergency call 1,<br />
emergency call 2, WC call, WC emergency call, cardiac alarm, telephone<br />
call, door call)?<br />
Are bed numbers (channel numbers) correctly presented?<br />
Can calls be answered at the console?<br />
Can a speech connection be established with the console via voice <strong>de</strong>vices<br />
(ComTerminal, patient handset)?<br />
Are presence states correctly displayed?<br />
Can announcements be broadcast from the console?<br />
NOTE ! If you note that a false bed number has been set at the connection<br />
socket combi bedhead unit, the medical supply unit must be opened in or<strong>de</strong>r<br />
to set the correct number (refer to documents on the medical supply unit).<br />
12.4 Checking functions between wards<br />
Check:<br />
Can ward coupling be activated at the ward console?<br />
Correct function of group lamps.<br />
76<br />