9. Activating the power supply<br />

9.2.3 Supply voltage to the room terminals<br />

Activate 5 calls and 5 presence states for the ward which is to be checked. Use<br />

the worst location, i.e. the most distant location from the power supply unit.<br />

Test the voltage and <strong>de</strong>termine the room terminal (ComTerminal <strong>Flamenco</strong>,<br />

RoomTerminal <strong>Flamenco</strong>, ControlTerminal <strong>Flamenco</strong> or ControlTerminal with<br />

doorplate <strong>Flamenco</strong>) with the lowest voltage. This point is consi<strong>de</strong>red as the<br />

electric centre of the ring circuit. Here, the voltage differential to the first room<br />

terminal as seen from the power supply unit must not be greater than 4 V.<br />

If you should read a lower value (i.e. a drop of more than 4 V) the voltage drop<br />

must be reduced by laying an additional cable from the power supply unit to the<br />

electric centre - or double the 0 V lead - if it is available.<br />

Power supply<br />

unit<br />

Ip<br />

Ip<br />

Electric<br />

centre<br />

9.2.4 Check voltage for group bus OSYnet<br />

If group bus OSYnet is correctly connected:<br />

voltage between DH and DG and<br />

voltage between DL and DG = approx. +2.5 V.<br />

9.2.5 Check voltage for RAN<br />

RAN voltage is o.k. when voltage between B (0V) and Y (RAN) is between<br />

+8.8 V and +9 V DC.<br />

If the voltage is outsi<strong>de</strong> of these parameters, these may be the reasons<br />

If voltage between B (0V) and Y (RAN) = 0 V<br />

a) Short circuit between B and Y<br />

b) Wire Y (RAN) is not connected to the connection socket.<br />

If voltage between B and Y = +12 V<br />

a) Wire B (0V) is not connected at one of the RAN users.<br />

b) One RAN user is <strong>de</strong>fective.<br />

62<br />