Microstream ® CO 2 Monitoring (Optional <strong>Passport</strong> 2) Operation<br />

Previous Menu<br />

Start<br />

CO2 Menu<br />

Calibration Menu<br />

Calculating, calibration gas can be removed.<br />

Date and Time of last successful Calibration<br />

3/15/99 at 11:44 AM<br />

Select to stop calibration<br />

7. After a moment, the message will change to “Calibration Completed Successfully”. The<br />

date and time of the successful calibration will appear in the body of the Calibration<br />

Menu.<br />

8. Rotate the Navigator Control Knob to the selection for “Previous Menu” and press to<br />

select.<br />

Previous Menu<br />

Start<br />

CO2 Menu<br />

Calibration Menu<br />

Calibration Completed Successfully<br />

Date and Time of last successful Calibration<br />

3/15/99 at 11:44 AM<br />

Select to stop calibration<br />

9. Rotate the Navigator Control Knob to select the “Normal Screen” and press to return to<br />

the monitors normal display screen.<br />

3 - 58 0070-10-0649-01 <strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions