Operation Gas Module (optional <strong>Passport</strong> 2)<br />
7. If not already set, use the Monitor Setup Menu to select the gas waveforms to be<br />
displayed in Waveform area 2, 3 or 4.<br />
8. If desired, the Gas waveform scale and speed can be changed by entering the Gases<br />
menu. See Section 2.3 for Menusdetails.<br />
NOTE: See <strong>Passport</strong> 2 / Gas Module Messages (only observed<br />
when Gas Module is installed) in Section 3.22.5 for more<br />
details on messages.<br />
3.18.2 Gas Module 3 Pre-use Test<br />
Prior to each use, perform the following test with the Gas Module 3 to verify that the gas<br />
analyzer and sample system are functioning properly:<br />
1. Verify that the appropriate water trap is properly installed and that the appropriate<br />
sampling line is connected.<br />
DRYLINE Adult/Pediatric water trap used with DRYLINE Adult/Pediatric sampling<br />
line (colorless Luer lock nut)<br />
DRYLINE Neonatal water trap used with DRYLINE Neonatal sampling line (blue<br />
Luer lock nut)<br />
2. Verify that the water trap container is less than half full.<br />
3. Occlude the sampling line and verify that the occlusion alarm functions properly.<br />
4. Breathe into the sampling line and verify that a CO2 waveform is correctly displayed on<br />
the monitor.<br />
5. Sample room air for 30 seconds and verify that the monitor oxygen output is 20.95%<br />
(± sensor inaccuracy).<br />
<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 61