Operation Status Messages<br />
3.22.3 Recorder Messages (only units equipped with recorder)<br />
MESSAGE REASON ACTION<br />
Local Printer Door Open The door of the printer is not<br />
closed.<br />
Close the printer door.<br />
Local Printer Out Of<br />
Paper<br />
The roll of printer paper is used<br />
up.<br />
Printer Busy Printer received multiple print<br />
requests at one time.<br />
Local Printer Unable To<br />
Print<br />
The system has detected an<br />
unrecoverable printer failure.<br />
3.22.4 CO 2 Messages (only units equipped with CO 2)<br />
MESSAGE REASON ACTION<br />
CO 2 : Filterline<br />
Disconnected<br />
CO 2: Warming<br />
Up<br />
CO 2 : Auto-zero In<br />
Progress<br />
CO 2 : Auto-zero<br />
Requested<br />
The filterline is not connected to the<br />
monitor.<br />
The CO 2 sensor has not reached its<br />
operating temperature. (The monitor<br />
was just turned on)<br />
The CO 2 sensor is performing an<br />
auto-zero.<br />
An Auto-Zero was automatically<br />
requested by the system.<br />
Replace with a new roll of paper.<br />
Wait until the printer is not busy.<br />
Power cycle unit. If message<br />
reappears, call <strong>Mindray</strong> DS<br />
Service.<br />
Connect the filterline.<br />
Wait for the message to go away. It<br />
takes typically 30 seconds for the<br />
sensor to warm up.<br />
Wait for the auto-zero to complete.<br />
Wait for the auto-zero to complete.<br />
CO 2 : Failure CO 2 system failure. Contact <strong>Mindray</strong> DS Technical<br />
Support.<br />
CO 2 : Occlusion The CO 2 hardware is indicating the<br />
sampling pump line is blocked while<br />
the CO 2 sidestream pump is on.<br />
CO 2: Purge The system has detected a blocked<br />
filterline and attempts to unblock it by<br />
temporarily increasing the flow rate.<br />
Check sampling line and filter for<br />
blockage, clear sampling line if<br />
possible. Replace sampling line if<br />
necessary.<br />
Disconnect and reconnect the filterline<br />
from the <strong>Passport</strong> 2 in order to<br />
clear this message.<br />
Check filterline and replace if<br />
necessary.<br />
3.22.5 <strong>Passport</strong> 2 / Gas Module Messages (only observed when Gas<br />
Module is installed)<br />
MESSAGE REASON ACTION<br />
GM: Warming Up Appears when the system has been<br />
turned on, and the sensors have not<br />
reached their stable operating<br />
temperature.<br />
GM: Agent<br />
Warming Up<br />
Does not apply to<br />
Gas Module 3.<br />
This message appears after the GMII:<br />
Warming Up message disappears. It<br />
indicates that the Agent ID Bench is<br />
warming up and readings will not be<br />
available.<br />
Wait for the message to go<br />
away. It takes up to two minutes<br />
for the device to warm up.<br />
Wait for the message to go<br />
away. It takes ups to five<br />
minutes from power-up for the<br />
Agent ID Bench to warm up.<br />
<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 81