Operation Status Messages<br />

3.22.3 Recorder Messages (only units equipped with recorder)<br />

MESSAGE REASON ACTION<br />

Local Printer Door Open The door of the printer is not<br />

closed.<br />

Close the printer door.<br />

Local Printer Out Of<br />

Paper<br />

The roll of printer paper is used<br />

up.<br />

Printer Busy Printer received multiple print<br />

requests at one time.<br />

Local Printer Unable To<br />

Print<br />

The system has detected an<br />

unrecoverable printer failure.<br />

3.22.4 CO 2 Messages (only units equipped with CO 2)<br />

MESSAGE REASON ACTION<br />

CO 2 : Filterline<br />

Disconnected<br />

CO 2: Warming<br />

Up<br />

CO 2 : Auto-zero In<br />

Progress<br />

CO 2 : Auto-zero<br />

Requested<br />

The filterline is not connected to the<br />

monitor.<br />

The CO 2 sensor has not reached its<br />

operating temperature. (The monitor<br />

was just turned on)<br />

The CO 2 sensor is performing an<br />

auto-zero.<br />

An Auto-Zero was automatically<br />

requested by the system.<br />

Replace with a new roll of paper.<br />

Wait until the printer is not busy.<br />

Power cycle unit. If message<br />

reappears, call <strong>Mindray</strong> DS<br />

Service.<br />

Connect the filterline.<br />

Wait for the message to go away. It<br />

takes typically 30 seconds for the<br />

sensor to warm up.<br />

Wait for the auto-zero to complete.<br />

Wait for the auto-zero to complete.<br />

CO 2 : Failure CO 2 system failure. Contact <strong>Mindray</strong> DS Technical<br />

Support.<br />

CO 2 : Occlusion The CO 2 hardware is indicating the<br />

sampling pump line is blocked while<br />

the CO 2 sidestream pump is on.<br />

CO 2: Purge The system has detected a blocked<br />

filterline and attempts to unblock it by<br />

temporarily increasing the flow rate.<br />

Check sampling line and filter for<br />

blockage, clear sampling line if<br />

possible. Replace sampling line if<br />

necessary.<br />

Disconnect and reconnect the filterline<br />

from the <strong>Passport</strong> 2 in order to<br />

clear this message.<br />

Check filterline and replace if<br />

necessary.<br />

3.22.5 <strong>Passport</strong> 2 / Gas Module Messages (only observed when Gas<br />

Module is installed)<br />

MESSAGE REASON ACTION<br />

GM: Warming Up Appears when the system has been<br />

turned on, and the sensors have not<br />

reached their stable operating<br />

temperature.<br />

GM: Agent<br />

Warming Up<br />

Does not apply to<br />

Gas Module 3.<br />

This message appears after the GMII:<br />

Warming Up message disappears. It<br />

indicates that the Agent ID Bench is<br />

warming up and readings will not be<br />

available.<br />

Wait for the message to go<br />

away. It takes up to two minutes<br />

for the device to warm up.<br />

Wait for the message to go<br />

away. It takes ups to five<br />

minutes from power-up for the<br />

Agent ID Bench to warm up.<br />