Operation SpO 2 Pulse Oximetry<br />
It is also possible to change the averaging time of the Saturation and Pulse Rate<br />
measurements. The post average time can be changed to 6, 8, 10, 12, 14 or 16 seconds.<br />
3.6.3 Performance Considerations<br />
3.6.3.1 Calibration<br />
3.6.3.2 Auto Scaling<br />
To ensure optimal performance, use an appropriate sensor, apply it as directed, and observe<br />
all warnings and cautions.<br />
If excessive ambient light is present, cover the sensor site with opaque material. Failure to do<br />
so may result in inaccurate measurements. Light sources that can affect performance include<br />
surgical lights, especially those with a xenon light source, bilirubin lamps, fluorescent lights,<br />
infrared heating lamps, and direct sunlight.<br />
In the event that you are unable to get any reading, or the reading you get is inaccurate,<br />
consider the following:<br />
If your patient is poorly perfused, try applying the sensor to another site - such as a<br />
different finger or toe.<br />
Check that the sensor is properly aligned.<br />
In electrosurgery, make sure sensor is not too close to ESU devices or cables.<br />
Check to make sure the site area is clean/non-greasy. Clean site and sensor if needed.<br />
Nail polish and fungus should be removed.<br />
The oximetry sub-system incorporates automatic calibration mechanisms. No other<br />
calibration is required.<br />
The Pleth waveform is automatically scaled. There is no adjustment that can be made to the<br />
Pleth waveform size.<br />
3.6.4 Masimo ® Sensors and Patient Cable<br />
Masimo provides a family of sensors suitable for a wide variety of clinical settings and<br />
patients. Specific sensors have been developed for neonates, infants, children, and adults.<br />
All sensors are:<br />
1. indicated for continuous non invasive monitoring of arterial oxygen saturation (SpO2 )<br />
and pulse rate<br />
2. non-sterile<br />
3. usable during patient movement<br />
The LNOP ® DCSC Adult Reusable Spot Check Sensor is used for “spot check” applications.<br />
The LNOP ® DCI Adult Re-usable Finger Sensor can also be used for “spot check”<br />
applications if needed. All sensors are intended for “single-patient use only” unless indicated<br />
as “reusable”.<br />
<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 29