Appendix Performance Specifications<br />

6.2.2.2 Arterial Blood Pressure<br />

(Only in monitors with serial number prefix “TS” and software versions AC.X and higher.)<br />

Bandwidth (-3 dB referenced to 10 Hz): DC to 15 Hz minimum.<br />

Propagation Delay: 25 ms, maximum.<br />

Sensitivity: 1 V/100 mmHg, ± 10%.<br />

6.2.2.3 Sync Pulse for Cardioversion<br />

Propagation Delay: 35 ms maximum, between QRS peak and the<br />

rising edge of the Sync Pulse.<br />

Amplitude: 2 V min.<br />

Width: 7 ms (3 Lead and 5 Lead ECG) 11 ms<br />

(12 Lead ECG)<br />

Source: The Sync Pulse is derived from the active ECG<br />

source.<br />

CAUTION: The Analog Output on the <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT<br />

supports triggering the Intra-Aortic Balloon Pump (IABP) for<br />

3 Lead and 5 Lead ECG cable monitoring only. Invasive<br />

Blood Pressure triggering is not supported. ECG analog<br />

output is disabled when 12 Lead ECG analysis is enabled.<br />

CAUTION: To assure successful triggering of Intra-Aortic balloon pump<br />

from the <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT monitor, set the “ECG<br />

Filter” to “Extended” and set “Pacer Enhancement” to<br />

“On”. Both of these settings are located in the ECG setup<br />

menu of the <strong>Passport</strong> 2/<strong>Passport</strong> 2 LT.<br />

6.2.3 Systole Detector and Heart Rate Meter<br />

The heart rate (HR) meter data source is automatically selected from the available monitored<br />

parameters. The selection priority is: (1) ECG; (2) IBP1; (3) IBP2; (4) SpO2; (5) NIBP<br />

6.2.3.1 ECG Derived Heart Rate Meter Performance Requirements<br />

Range: 30 to 300 bpm Adult / Pediatric<br />

Resolution: 1 bpm.<br />

30 to 350 bpm Neonatal<br />

(3-Lead or 5-Lead ECG)<br />

30 to 300 bpm Neonatal (12-Lead ECG)<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 7