Appendix Performance Specifications<br />

6.2.6.5 Maximum Leakage<br />

6.2.6.6 Vent Rate<br />

The maximum allowed pressure drop with the bleed valves closed is 10 mmHg in 10 seconds<br />

as measured with a 500 cc volume at differential pressures of 250 mmHg, 150 mmHg and<br />

50 mmHg (reference ANSI/AAMI SP10-1992, EN1060-3, 1997).<br />

A volume of 500 cc, when vented, is reduced from a pressure of 260 mmHg to a pressure of<br />

15 mmHg in a maximum of 10 seconds. For Neonate, from a pressure of 150 mmHg to a<br />

pressure of 5 mmHg in less then 5 seconds (reference: EN1060-3, 1997).<br />

6.2.7 NIBP Sub-System Functional Requirements<br />

6.2.7.1 Initial Conditions<br />

The NIBP System performs a zero calibration adjustment at power-up.<br />

6.2.7.2 NIBP Starting Pressure Settings and Ranges<br />

Range of Adjustment: Adult Patients, 100-280 mmHg<br />

6.2.7.3 NIBP Measurement Cycle<br />

Pediatric Patients, 60-180 mmHg<br />

Neonatal Patients, 40-120 mmHg<br />

Default Start Pressure: Adult Patients, 180 mmHg.<br />

Pediatric Patients, 140 mmHg.<br />

Neonatal Patients, 100 mmHg.<br />

Pressure Increment: 5 mmHg, regardless of patient size.<br />

There are two different modes of measurement operation: manual and interval modes. The<br />

manual mode requires the operator to initiate the measurement cycle. The interval mode<br />

follows a configured plan of automatically initiated measurement cycles.<br />

The Maximum Measurement Cycle Duration shall be 180 sec for Adult and Pediatric patients<br />

(reference EN601-2-30, 1995).<br />

The Maximum Measurement Cycle Duration shall be 90 sec for Neonatal patients (reference<br />

EN601-2-30, 1995).<br />

During a measurement, if the initial cuff inflation pressure is found to be inadequate, the unit<br />

shall retry with a higher inflation pressure (+50 ±10 mmHg in Adult and Pediatric modes and<br />

+40 ±10 mmHg in Neonatal mode).<br />

In interval mode only, the unit shall adjust the inflation pressure according to the previous<br />

systolic pressure. After the first successful measurement is made, the subsequent inflation<br />

pressure becomes +50 ±10 mmHg in the Adult and Pediatric modes and +40 ±10 mmHg in<br />

Neonatal mode.<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 13