6.0 Appendix<br />

6.1 Safety Designations<br />

6.1.1 Safety designations per IEC 60601-1 Standard<br />

Type of protection against electrical shock: Class 1 and Internal Electric Power Source.<br />

Where the integrity of the external protective<br />

earth in the installation or its conductor<br />

arrangement is in doubt, equipment shall be<br />

operated from its internal electric power<br />

source.<br />

Degree of protection against electric shock: ECG and IBP, Type CF defibrillation protected.<br />

NIBP, Type BF defibrillation protected. SpO2 ,<br />

CO2 , and Temperature, Type BF<br />

Supply Connection: 100V - 240VAC<br />

50 or 60 Hz.<br />

1.2 - 0.7 A<br />

12 VDC Sealed Lead Acid Internal Battery<br />

or<br />

11.1 VDC Lithium-Ion Internal Battery<br />

Mode of Operation: Continuous<br />

Protection Against Hazards of Explosion: Not Protected (Ordinary)<br />

Protection Against Ingress of Liquids: Not Protected (Ordinary)<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 1