Performance Specifications Appendix<br />
CMRR: 90 dB min., Maximum, output of 1mV p-p (RTI)<br />
over a 60 second period at 50/60 Hz, with a<br />
parallel combination 51KΩ and 0.047uF<br />
imbalance and ± 300 mV DC offset per AAMI<br />
EC13-1992 3.2.9.10 or ANSI/AAMI EC 11-<br />
1991,3.2.11.<br />
Lead Fault: Lead resistance ≤ 51KΩ in parallel with<br />
0.047µF shall not cause a lead fault<br />
condition. Also, a ± 300 mV offset shall not<br />
cause a lead fault condition.<br />
Pacer Enhancement: Pacer signals with amplitudes within the range<br />
± 2mV and ± 700 mV (RTI) amplitude with a<br />
maximum rise time of 100usec and with<br />
duration in the 0.1ms to 2.0ms range shall be<br />
enhanced when the Pacer Mode is turned<br />
ON.<br />
ESIS Filtering:<br />
(3 Lead and 5 Lead ECG configuration):<br />
6.2.2 Analog Output Specifications<br />
6.2.2.1 3 Lead and 5 Lead ECG<br />
Greater than 90 dB attenuation at 500 kHz.<br />
ESU Noise Detection: An ESU noise declaration will be asserted if<br />
signals detected as pacers are detected at a<br />
rate greater than 50 Hz. No noise declaration<br />
will be asserted if pacer-like signals are<br />
detected at a rate less than 10 Hz. Applies to<br />
3 Lead and 5 Lead ECG configurations only.<br />
Propagation Delay (Delay of QRS complex): 25 ms maximum.<br />
Sensitivity (referenced to 10Hz: 1 V/mV of input, ± 10%.<br />
Pacer Enhancement: Pacer is summed at the output when pacer<br />
enhance mode is turned ON. Pacer Amplitude<br />
signal is a minimum of 2.5V Pacer width is 10<br />
ms, with a 5% tolerance. Pacer rise and fall<br />
times are 100 ms, maximum.<br />
6 - 6 0070-10-0649-01 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions