Appendix Performance Specifications<br />

Pacemaker Pulse Rejection: Rejects pulses of amplitude ± 2.0 mV to ± 700<br />

mV and duration 0.1 ms to 2 ms with no tail<br />

(ANSI/AAMI EC13-1992, 3.1.4.1).<br />

6.2.3.2 IBP Derived Heart Rate Meter Performance<br />

Refer to Section 6.2.8.4<br />

6.2.3.3 SpO 2 Derived Heart Rate Meter Performance<br />

Refer to Section 6.2.11.2 and 6.2.11.4<br />

6.2.3.4 NIBP Derived Heart Rate Meter<br />

Refer to Section 6.2.6.2<br />

6.2.4 S-T Segment Analysis<br />

3/5 Lead ECG:<br />

6.2.4.1 S-T Segment Analysis Performance Requirements<br />

Rejects all pacer pulses ± 2.0 mV to ± 700<br />

mV and duration 0.1 ms to 2 ms with 4 ms<br />

time constant tail of less than 2.0 mV.<br />

12 Lead ECG:<br />

Does not reject pacer pulses with tails<br />

(overshoot/undershoot).<br />

Trigger Indication: Audible beep on every beat captured.<br />

Heart Rate Averaging: 3/5/12 Lead ECG: The average heart rate is<br />

calculated as follows:<br />

Mean R to R interval in the last 16 R to R<br />

intervals (HR > 48 bpm).<br />

Mean R to R interval in the last 4 R to R<br />

intervals (HR ≤ 48 bpm).<br />

Enabling: Available in ADULT and PEDIATRIC modes<br />

only.<br />

ST Deviation Range:<br />

(3-Lead and 5-Lead ECG)<br />

9.9 mm to + 9.9mm<br />

(-990µV to + 990µV RTI)<br />

Resolution: 0.1 mm (10 µV)<br />

Default ST Measurement Point: 80 ms after J point for heart rates ≤ 120 bpm.<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 6 - 9