Operation Gas Module (optional <strong>Passport</strong> 2)<br />

3.18 Gas Module (optional <strong>Passport</strong> 2)<br />

The Gas Module option allows for the measurement of anesthetic gases, O2, N2O and CO2 levels. Measurement can be acquired via a nasal cannula (non-intubated) for oxygen and<br />

CO2 only, or through a sampling line connected to a breathing circuit (intubated). For a<br />

complete listing of accessories, refer to Chapter 5.0.<br />

NOTE: The <strong>Passport</strong> 2 will interface to the Gas Module via the SP1<br />

Connector on the Comm-Port mounted onto the <strong>Passport</strong> 2.<br />

WARNING: When using the Gas Module, the maximum sampling rate at<br />

the nasal cannula is 200 ml/min (120 ml/min for Gas<br />

Module 3 with a neonatal water trap). This device should<br />

not be used on patients whose breathing could be impaired<br />

by this vacuum flow rate.<br />

NOTE: The Gas Module 3 is equipped with automatic barometric<br />

pressure compensation.<br />

NOTE: The Gas Module 3 uses a fixed correction of 11 hPa to<br />

compensate for the influence of water vapor in the gas<br />

sample, when converting the gas readings to ATPD. An<br />

increase in the ambient H 20 partial pressure to 30 hPa<br />

(i.e. 28 °C, 80% RH or 33 °C, and 60% RH) will cause a<br />

general error for all gases of only -2% REL.<br />

3.18.1 Sequence for Monitoring Anesthetic Gases, O 2 , N 2 O and/or CO 2<br />

NOTE: To prevent moisture from entering the pneumatic system,<br />

ensure that the Gas Module is always installed and<br />

operated in the horizontal orientation shown in all graphical<br />

depictions.<br />

1. Turn on the Gas Module and <strong>Passport</strong> 2, and configure the <strong>Passport</strong> 2 Serial Port to<br />

be used with the Gas Module. Set alarms as necessary.<br />

7 Foot Nasal Cannula, CO 2<br />

P/N 0683-00-0424-10<br />

Adapter, Straight Tee ET<br />

P/N 0683-00-0242-22<br />

7 Foot Nasal Cannula, CO 2 /O 2<br />

P/N 0683-00-0452-10<br />

FIGURE 3-24 Gas Module II and SE Airway Adapter<br />

®<br />

<strong>Passport</strong> 2<br />

Patient/ADT Monitor Setup Alarms<br />

Trend<br />

Record<br />

11 /9/98 10 :28 am<br />

This is app roxi matl y whereth e top l evel info. wou ld app ear Juan Valdez /Roo m 1420 / BED 2<br />

HR<br />

Lead mm<br />

P<br />

II + 2.2<br />

S<br />

aV R -2.1<br />

D<br />

PVC/min: 45 V1 -3.0<br />

A Sou rce: Auto ST Pt: 80 /60<br />

RPM<br />

R<br />

E<br />

S 18<br />

P<br />

Sou rc e : ECG<br />

<strong>Passport</strong> 2 ® /<strong>Passport</strong> 2 LT Operating Instructions 0070-10-0649-01 3 - 59<br />

8<br />

T1<br />

T<br />

E<br />

M<br />

P<br />

36.7C<br />

10 Foot Sample Line, Patient<br />

P/N 0683-00-0451-10<br />

Adapter, Mask Elbow ET<br />

P/N 0683-00-0242-12<br />

A<br />

G O2<br />

T<br />

Des<br />

1 N 2O Ins p Exp<br />

C<br />

0<br />

2 1 /38mm<br />

Hg<br />

RR (CO2 ) 14 rp m<br />

Sp O2<br />

PR<br />

S<br />

P<br />

0<br />

2<br />

% 81bp<br />

m<br />

In sp Exp<br />

Sy s<br />

Dia<br />

N<br />

41 36 % I<br />

B 120/ 80<br />

7.6 5.2%<br />

P Mean (93)<br />

51 54 %<br />

Interval: ON ET: 13 min<br />

or<br />

Gas Scavenging Adapter Assy.<br />

P/N 0997-00-0923<br />

or<br />

P/N 0997-00-0984